Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-2185-2016

Target Detachable Coils: TARGET HELICAL NANO 3MM X 4CM MODEL Number:M0035453040 Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.

Publicado el 20 de julio de 2016 · Inicio del retiro: 6 de junio de 2016

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Stryker Neurovascular
Distribuidor
Worldwide Distribution - US (nationwide) and Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italia, Japan,Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Mexico,Netherlands, Norway, Peru, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom, United States, Uruguay, and Vietnam.
Lote
MODEL Number:M0035453040; Lot Number/Expiration Date: 18603553 30-Sep-18
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta