FDAFDA · Alimentos y medicamentos·N.º oficial Z-2195-2013
AOA Mini RPE Screw Assembly , a component of Herbst, MARA and Fixed Metal devices. Product Usage: AOA Mini RPE Screw Assembly is a component of Herbst, MARA and Fixed Metal devices; removable functional appliances that are to modify the growth of the jaws in a prescribed growth pattern, however, this category is generally attached to the patients upper and lower molars or bicuspids by means of temporary stainless steel crowns or orthodontic bands. Retainer, Screw Expansion, Orthodontic.
Publicado el 18 de septiembre de 2013 · Inicio del retiro: 6 de junio de 2013
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The firm initiated the recall of AOA Mini RPE Screw Assembly, because it may be missing a complete weld joining one side of the arm with the screw body. Herbst, MARA and Fixed Metal devices may be affected, because they contain an AOA Mini RPE Screw Assembly.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Sybron Dental Specialties
- Distribuidor
- Worldwide Distribution - USA Nationwide and the countries of Australia, Canada, and EU.
- Lote
- The AOA Mini RPE Screw Assembly is a component used in customized palate expander appliances assembled by prescription. The affected assembly may be found in products listed in the table below: Product Description: Herbst Bite Jumper Herbst Appliance Acrylic Splint M.A.R.A. (Mandibular Anterior Reposi tioning Appliance) Rapid Palatal Expander Lower Expander DeLuke Bonded Frameless Expander Haas Expander AOA Mini Rapid Palatal Expander Dischinger Lower Expansion Appliance Tracy/Hilgers MDA Appliance Distal Jet Molar Uprighter Part Number: 3000-0010 3000-0011 3000-0012 3000-0040 3000-0042 3000-0045 3000-0046 3000-0047 3000-0050 3000-0094 3000-0097 Lot Number: "All Devices Manufactured from April 3, 2013 to May 21, 2013"
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States