LVIS Intraluminal Support Device, REF/Outside Diameter(OD)xTotal Length(TL)xWorking Length(WL): 214022-CAS/4.0 mm x 22 mm x 18 mm, 214518-CAS/4.5 mm x 18 mm x 14 mm, 213517-CAS/3.5 mm x 17 mm x 13 mm
Publicado el 20 de mayo de 2026 · Inicio del retiro: 6 de abril de 2026
Resumen
FDA registra el retiro Z-2200-2026 de dispositivos médicos de Clase II, publicado el 20 de mayo de 2026 e iniciado el 6 de abril de 2026. Empresa responsable: MICROVENTION INC.; Estado: En curso; Cantidad reportada: 38; Lote: Lots: 0000616462, 0000619851, 0000552987, 0000551725. UDI: (01)00842429115893(11)240325(17)270228(10)0000552987, (01)00842429115893(11)240325(17)270228(10)0000552987, (01)00842429115893(11)240325(17)270228(10)0000552987, (01)00842429115930(11)240325(17)270228(10)0000551725, (01)00842429115930(11)240325(17)270228(10)0000551725, (01)00842429115930(11)240610(17)270531(10)0000616462, (01)00842429115930(11)240610(17)270531(10)0000616462, (01)00842429115961(11)240610(17)270531(10)0000619851, (01)00842429115961(11)240610(17)270531(10)0000619851, (01)00842429115961(11)240610(17)270531(10)0000619851, (01)00842429115961(11)240610(17)270531(10)0000619851,. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Stent-assisted coiling device may due to commingling during manufacturing have dimensional nonconformance and/or incorrect device configuration. If device is oversized may lead to increased vessel wall stress, neointimal hyperplasia, vessel injury/rupture; if undersized may lead to incomplete expansion/malposition, acute stent thrombosis, distal embolization, ischemic neurological events
Qué hacer
Voluntary: Firm initiated
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- MICROVENTION INC.
- Distribuidor
- International distribution to the country of China.
- Lote
- Lots: 0000616462, 0000619851, 0000552987, 0000551725. UDI: (01)00842429115893(11)240325(17)270228(10)0000552987, (01)00842429115893(11)240325(17)270228(10)0000552987, (01)00842429115893(11)240325(17)270228(10)0000552987, (01)00842429115930(11)240325(17)270228(10)0000551725, (01)00842429115930(11)240325(17)270228(10)0000551725, (01)00842429115930(11)240610(17)270531(10)0000616462, (01)00842429115930(11)240610(17)270531(10)0000616462, (01)00842429115961(11)240610(17)270531(10)0000619851, (01)00842429115961(11)240610(17)270531(10)0000619851, (01)00842429115961(11)240610(17)270531(10)0000619851, (01)00842429115961(11)240610(17)270531(10)0000619851,
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDABOBBY Balloon Guide Catheter REF: BOB-895-CN BOB-895-EMEA BOB-895-EMEA-AL BOB-895-US MV-BGA01095 For use in19 de agosto de 2026
- FDAMedtronic Marinr Steerable, Product Number 72402; REPROCESSED ELECTROPHYSIOLOGY CATHETER20 de mayo de 2026
- FDAMedtronic SynchroMed Flex Infusion Mode - A810 Clinician Programmer Software Application, loaded on the20 de mayo de 2026
- FDAMicrovention, Headway" 17 Advanced Straight Microcatheter with Hydrophilic Coating, REF: MC172150SX-CN,29 de enero de 2025
- FDAERIC Retrieval Device REF: ER173020 ER174030 ER176044 The ERIC Retrieval Device is a mechanical thrombectomy25 de septiembre de 2024
- FDATERUMO HydroPearl Compressible Microspheres for Embolisation REF 8HP2S6007 de febrero de 2024
Preguntas sobre este retiro
¿LVIS Intraluminal Support Device, REF/Outside Diameter(OD)xTotal Length(TL)xWorking Length(WL): 214022-CAS/4.0 mm x 22 mm x 18 mm, 214518-CAS/4.5 mm x 18 mm x 14 mm, 213517-CAS/3.5 mm x 17 mm x 13 mm fue retirado del mercado?
Sí. FDA publicó el retiro Z-2200-2026 el 20 de mayo de 2026. Peligro reportado: Stent-assisted coiling device may due to commingling during manufacturing have dimensional nonconformance and/or incorrect device configuration. If device is oversized may lead to increased vessel wall stress, neointimal hyperplasia, vessel injury/rupture; if undersized may lead to incomplete expansion/malposition, acute stent thrombosis, distal embolization, ischemic neurological events
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated