Brilliance CT 6, M/N 728256; Brilliance CT 16 M/N 728246; Brilliance CT 40, M/N 728235 & Brilliance CT 64 M/N 728231, Philips Healthcare, Cleveland, OH 44143. Intended to produce cross-sectional images of the body by computer reconstruction.
Publicado el 25 de septiembre de 2013 · Inicio del retiro: 5 de julio de 2013
Resumen
FDA registra el retiro Z-2210-2013 de dispositivos médicos de Clase II, publicado el 25 de septiembre de 2013 e iniciado el 5 de julio de 2013. Empresa responsable: Philips Medical Systems (Cleveland) Inc; Estado: Finalizado; Cantidad reportada: M/N: 728256, 4 units; M/N: 728246, 89 units; M/N: 728235, 9 units & M/N: 728231, 207 units; Lote: Model #, 728231, Serial #: 4021, 9535, 9544, 9587, 9627, 9701, 9707, 9713, 9759, 9875, 9887, 9917, 9925, 9931, 9980, 9996, 10009, 10011, 10012, 10015, 10028, 10036, 10042, 10045, 10050, 10052, 10062, 10063, 10064, 10077, 10079, 10081, 10096, 10099, 10102, 10111, 10114, 10120, 10122, 10135, 10136, 10137, 10139, 10145, 10148, 10149, 10151, 10152, 10153, 10154, 10162, 10165, 10166, 10170, 10171, 10172, 10173, 10188, 10189, 10191, 10196, 10198, 10203, 10214, 10222, 10227, 10228, 10229, 10230, 10235, 10238, 10242, 10243, 10247, 10259, 10260, 10261, 10263, 10268, 10273, 10274, 10278, 10288, 10291, 10293, 10294, 10297, 10298, 10301, 10302, 10307, 10312, 10313, 10316, 10317, 10318, 10319, 10320, 10327, 10336, 10337, 10341, 10347, 10349, 10350, 10351, 10352, 10353, 10362, 10365, 10376, 10380, 10384, 10387, 10393, 10400, 10411, 10412, 10414, 10431, 10438, 10439, 10445, 10446, 10452, 10453, 10459, 10461, 10468, 10470, 10471, 10476, 10482, 10486, 10490, 10491, 10503, 10511, 10513, 10519, 10535, 10536, 10540, 10544, 10546, 10551, 10554, 10561, 10563, 10568, 10569, 10574, 10582, 10600, 10617, 10628, 10631, 10635, 10646, 10647, 10651, 10655, 10659, 10673, 10684, 10686, 10698, 10714, 10733, 10734, 29002, 29014, 29016, 29027, 90054, 90099, 95022, 95046, 95073, 95103, 95107, 95109, 95114, 95132, 95133, 95181, 95209, 95268, 95295, 95305, 95371, 95453, 95473, 95563, 95660, 95704, 95706, 95707, 95712, 95715, 95721, 95727, 95729, 95769, 95778, 90199 & 95162. Model #, 728235, Serial #: 9034, 9190, 9233, 9234, 9261, 29049, 29057, 29064 & 29149. Model #, 728246, Serial #: 3150, 3163, 4051, 5547, 5585, 5623, 5655, 5666, 5670, 5715, 5724, 5747, 5773, 5776, 5791, 5803, 5813, 5836, 5855, 5862, 5868, 5876, 5877, 5897, 5911, 5916, 5935, 5969, 5971, 5974, 5997, 6001, 6006, 6008, 6028, 6087, 6102, 6117, 6120, 6121, 6126, 6170, 6195, 6200, 6202, 6203, 6238, 6243, 6281, 6287, 6306, 6318, 6324, 6346, 6352, 6387, 6402, 6405, 6444, 6484, 6491, 6494, 6495, 6496, 6503, 6517, 6519, 6523, 6556, 6559, 6609, 6610, 6611, 6612, 6623, 6624, 6640, 6642, 50115, 50147, 50166, 50180, 50189, 50193, 50195, 50197, 50198, 50204 & 50043. Model #, 728256, Serial #: 3490, 3500, 3679 & 3693. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The patient support may move in an unintended manner if the footswitch cover impinges on the footswitch.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Philips Medical Systems (Cleveland) Inc
- Distribuidor
- Worldwide Distribution-USA (nationwide) including DC and the states of AL, AR, AZ, CA, CO, DE, FL, GA, IN, KY, MA, MD, MI, MN, MS, NC, NM, NV, NY, OH, OK, PA, SC, TN, TX, VT, WI & WV, and the countries of Australia, Austria, Bosnia & Herzegovina, Chile, China, Croatia, Denmark, Equatorial Guinea, France, Germany, Greenland, Iceland, Indonesia, Israel, Japan, Kazakhstan, Netherlands, Nigeria, Norway, Poland, Romania, South Africa, South Korea, Sweden, Switzerland, Taiwan, United Arab Emirates, United Kingdom & Vietnam.
- Lote
- Model #, 728231, Serial #: 4021, 9535, 9544, 9587, 9627, 9701, 9707, 9713, 9759, 9875, 9887, 9917, 9925, 9931, 9980, 9996, 10009, 10011, 10012, 10015, 10028, 10036, 10042, 10045, 10050, 10052, 10062, 10063, 10064, 10077, 10079, 10081, 10096, 10099, 10102, 10111, 10114, 10120, 10122, 10135, 10136, 10137, 10139, 10145, 10148, 10149, 10151, 10152, 10153, 10154, 10162, 10165, 10166, 10170, 10171, 10172, 10173, 10188, 10189, 10191, 10196, 10198, 10203, 10214, 10222, 10227, 10228, 10229, 10230, 10235, 10238, 10242, 10243, 10247, 10259, 10260, 10261, 10263, 10268, 10273, 10274, 10278, 10288, 10291, 10293, 10294, 10297, 10298, 10301, 10302, 10307, 10312, 10313, 10316, 10317, 10318, 10319, 10320, 10327, 10336, 10337, 10341, 10347, 10349, 10350, 10351, 10352, 10353, 10362, 10365, 10376, 10380, 10384, 10387, 10393, 10400, 10411, 10412, 10414, 10431, 10438, 10439, 10445, 10446, 10452, 10453, 10459, 10461, 10468, 10470, 10471, 10476, 10482, 10486, 10490, 10491, 10503, 10511, 10513, 10519, 10535, 10536, 10540, 10544, 10546, 10551, 10554, 10561, 10563, 10568, 10569, 10574, 10582, 10600, 10617, 10628, 10631, 10635, 10646, 10647, 10651, 10655, 10659, 10673, 10684, 10686, 10698, 10714, 10733, 10734, 29002, 29014, 29016, 29027, 90054, 90099, 95022, 95046, 95073, 95103, 95107, 95109, 95114, 95132, 95133, 95181, 95209, 95268, 95295, 95305, 95371, 95453, 95473, 95563, 95660, 95704, 95706, 95707, 95712, 95715, 95721, 95727, 95729, 95769, 95778, 90199 & 95162. Model #, 728235, Serial #: 9034, 9190, 9233, 9234, 9261, 29049, 29057, 29064 & 29149. Model #, 728246, Serial #: 3150, 3163, 4051, 5547, 5585, 5623, 5655, 5666, 5670, 5715, 5724, 5747, 5773, 5776, 5791, 5803, 5813, 5836, 5855, 5862, 5868, 5876, 5877, 5897, 5911, 5916, 5935, 5969, 5971, 5974, 5997, 6001, 6006, 6008, 6028, 6087, 6102, 6117, 6120, 6121, 6126, 6170, 6195, 6200, 6202, 6203, 6238, 6243, 6281, 6287, 6306, 6318, 6324, 6346, 6352, 6387, 6402, 6405, 6444, 6484, 6491, 6494, 6495, 6496, 6503, 6517, 6519, 6523, 6556, 6559, 6609, 6610, 6611, 6612, 6623, 6624, 6640, 6642, 50115, 50147, 50166, 50180, 50189, 50193, 50195, 50197, 50198, 50204 & 50043. Model #, 728256, Serial #: 3490, 3500, 3679 & 3693.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAPinnacle3 Radiation Therapy Planning System, Model numbers 870231 and 870237.21 de septiembre de 2022
- FDABrilliance 64 Model # 728231, computed tomography x-ray system14 de agosto de 2019
- FDAIngenuity Core128 Model # 728323, computed tomography x-ray system14 de agosto de 2019
- FDABrightView XCT , Model Number 88248231 de julio de 2019
- FDABlanketrol III, Model Number 233, Cincinnati Sub-Zero Products, Inc., 12011 Mosteller Road, Cincinnati, OH25 de septiembre de 2013
- FDADevice Name: Terino Square Chin-Style II-Medium; Catalog Number: TSCII-M. To augment or reconstruct25 de septiembre de 2013
Preguntas sobre este retiro
¿Brilliance CT 6, M/N 728256; Brilliance CT 16 M/N 728246; Brilliance CT 40, M/N 728235 & Brilliance CT 64 M/N 728231, Philips Healthcare, Cleveland, OH 44143. Intended to produce cross-sectional images of the body by computer reconstruction. fue retirado del mercado?
Sí. FDA publicó el retiro Z-2210-2013 el 25 de septiembre de 2013. Peligro reportado: The patient support may move in an unintended manner if the footswitch cover impinges on the footswitch.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter