DIAMOND-FLEX CIRCULAR RETRACTOR, ANGLED, 40MM 5MM, Product Code 89-6114, Lot Code(s): 878971; 879366; 879404, and Date Code: C16. Designed to retract or elevate organs and tissue to provide better visualization access.

Publicado el 3 de agosto de 2016 · Inicio del retiro: 1 de julio de 2016

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

BD, formerly CareFusion, has identified a potential risk associated with a weld failure which could result in the wire protruding thru the tip of the instrument when articulated. If this failure were to occur while in use in a procedure it has the potential to damages tissue or organs

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Carefusion 2200 Inc
Distribuidor
Worldwide Distribution - US, including the states of WA and MD; and, the country of Japan.
Lote
Product Code 89-6114, Lot Code(s): 878971; 879366; 879404, and Date Code: C16
Clasificación de riesgo
Class I