Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-2220-2012

Altrx" Altralinked" Polyethylene liners Description: An acetabular liner is one of the implants used during total hip arthroplasty to replace the damaged hip joint to help increase patient mobility and to reduce pain. Packaging: Product is placed into a PETG tray, sealed with a Tyvek lid. This is placed into a second PETG tray and sealed with a Tyvek lid. This package is placed into a paperboard carton, labeled, and shrink-wrapped. Intended Use: An acetabular liner is used in combination with other implants for total hip arthroplasty is intended to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components.

Publicado el 29 de agosto de 2012 · Inicio del retiro: 24 de julio de 2012

Resumen

FDA registra el retiro Z-2220-2012 de dispositivos médicos de Clase II, publicado el 29 de agosto de 2012 e iniciado el 24 de julio de 2012. Empresa responsable: DePuy Orthopaedics, Inc.; Estado: Finalizado; Cantidad reportada: 20 liners; Lote: Catalog number: 1221-36-054 and Lot Number: 231859. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

In the recalled lot, the products were not machined according to the production router, a size 52 cup (122136052) was machined instead. This resulted in a size 52 liner being etched and labeled as a size 54 liner.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
DePuy Orthopaedics, Inc.
Distribuidor
Worldwide Distribution - USA including PA and the country of Canada
Lote
Catalog number: 1221-36-054 and Lot Number: 231859
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Altrx" Altralinked" Polyethylene liners Description: An acetabular liner is one of the implants used during total hip arthroplasty to replace the damaged hip joint to help increase patient mobility and to reduce pain. Packaging: Product is placed into a PETG tray, sealed with a Tyvek lid. This is placed into a second PETG tray and sealed with a Tyvek lid. This package is placed into a paperboard carton, labeled, and shrink-wrapped. Intended Use: An acetabular liner is used in combination with other implants for total hip arthroplasty is intended to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components. fue retirado del mercado?

Sí. FDA publicó el retiro Z-2220-2012 el 29 de agosto de 2012. Peligro reportado: In the recalled lot, the products were not machined according to the production router, a size 52 cup (122136052) was machined instead. This resulted in a size 52 liner being etched and labeled as a size 54 liner.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter