Hitachi Echelon MRI System/Oasis MRI, Software. These are software controlled MRI devices. Product Usage: The Echelon and Oasis MR systems are imaging devices, intended to provide the physician with physiological and clinical information that can be useful in diagnosis determination.

Publicado el 29 de agosto de 2012 · Inicio del retiro: 15 de junio de 2012

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Hitachi discovered a software error with a feature called Multi-Planar Reconstruction (MPR). MPR takes an original volume dataset reconstructed in any plane view and creates a new image set from another view (i.e. axial to coronal, coronal to sagittal, etc.). The error causes the first image in the dataset to have a right/left reversed orientation. The software error can occur randomly when execu

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Hitachi Medical Systems America Inc
Distribuidor
US Nationwide Distribution - including the states of:: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WY and Mexico.
Lote
Echelon, V001 through V061, inclusive. Oasis, M001 through M105, inclusive, M951
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States