DRI Cocaine Metabolite Assay, labeled as Multigent Cocaine Assay for Abbott Laboratories. 100 ml, IVD Mfg by: Microgenics Corporation, 46360 Fremont Blvd. Fremont, CA 94538 The DRI Cocaine Metabolite Enzyme Immunoassay is a homogeneous enzyme intended for the qualitative and semi-quantitative determination of benzoylecgonine (Cocaine Metabolite) in human urine with either 300 ng/mL or 150 ng/mL as a cutoff calibrator. The assay provided only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirm analytical result.

Publicado el 2 de octubre de 2013 · Inicio del retiro: 23 de julio de 2013

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Surveillance of reagent stability for specific lots of DRI Cocaine Metabolite Reagent show an unexpected decreasing change in the reaction rate, indicating a faster than expected reagent degradation.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Microgenics Corp
Distribuidor
US Distribution including the states of DC, MA, NJ, TX. FL, CA, WA, NY, TX, CT, KY, LA, AR, TN, AL, GA, CT, VA, NC, KS, MD, OH and IL
Lote
Part no. 0055, lot number 60053405, Exp 2/29/2016; Part No. 0056, lot number 60053396, Exp 2/29/2016; Part No. 10014593, lot number 60112588, Exp 2/29/2016; - not distributed Part number #l40-20, lot number 24215M500, Exp 8/13/14.
Clasificación de riesgo
Class II