TRESTLE LUXE® Anterior Cervical Plating System, Product code KWQ. 510(k) K102820. The TRESTLE LUXE Anterior Cervical Plating System is a temporary device used to stabilize the cervical spine during bone fusion development. Device implants include a range of plate sizes and bone screws to provide the versatility required for the specific indications noted. Fixation is achieved by means of a rigid plate that is surgically attached to the spine with bone screws. The TRESTLE LUXE Anterior Cervical Plating System is intended for use in the anterior cervical spine (C2-C7). Patients with trauma (including fractures), spondylolisthesis, and pseudoarthrosis, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spinal stenosis, tumors, and failed previous fusion.
Publicado el 5 de septiembre de 2012 · Inicio del retiro: 29 de mayo de 2012
Resumen
FDA registra el retiro Z-2273-2012 de dispositivos médicos de Clase II, publicado el 5 de septiembre de 2012 e iniciado el 29 de mayo de 2012. Empresa responsable: Alphatec Spine, Inc.; Estado: Finalizado; Cantidad reportada: 38; Lote: Product code KWQ. 510(k) K102820 Part Number 71721. Lot numbers: 6629701, 6594701, 6434301, 6434302, 6434302R1. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
On May 9th 2012 Alphatec Spine received a complaint which reported that the distal tip of the Trestle Luxe Slide Alignment Tool (Part Number 71721), does not properly fit into the hexalobe screw on the Trestle Luxe Plate.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Alphatec Spine, Inc.
- Distribuidor
- Nationwide Distribution.
- Lote
- Product code KWQ. 510(k) K102820 Part Number 71721. Lot numbers: 6629701, 6594701, 6434301, 6434302, 6434302R1.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDASafeOp Patient Interface, REF: AIX11240, part of SafeOp 3: Neural Informatix System29 de julio de 2026
- FDAIdentiTi ALIF Standalone Interbody System: Part Number/Description 132-10-85-250 / IdentiTi ALIF SA Graft6 de marzo de 2024
- FDAAtec Insignia Anterior Cervical Plate System, REF 136-0354 Insignia, ACP, 3-Level, 54 mm, Rx Only,16 de junio de 2021
- FDAAtec Insignia Anterior Cervical Plate System, REF 136-0248 Insignia, ACP, 2-Level, 48 mm, Rx Only,16 de junio de 2021
- FDASunquest Laboratory LabAccess Results Workstation (LARS)5 de septiembre de 2012
- FDAApelo Modular Screwdriver labeled in part: "APELO PEDICLE SCREW SYSTEM***Modular Screwdriver***Atlas Spine,5 de septiembre de 2012
Preguntas sobre este retiro
¿TRESTLE LUXE® Anterior Cervical Plating System, Product code KWQ. 510(k) K102820. The TRESTLE LUXE Anterior Cervical Plating System is a temporary device used to stabilize the cervical spine during bone fusion development. Device implants include a range of plate sizes and bone screws to provide the versatility required for the specific indications noted. Fixation is achieved by means of a rigid plate that is surgically attached to the spine with bone screws. The TRESTLE LUXE Anterior Cervical Plating System is intended for use in the anterior cervical spine (C2-C7). Patients with trauma (including fractures), spondylolisthesis, and pseudoarthrosis, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spinal stenosis, tumors, and failed previous fusion. fue retirado del mercado?
Sí. FDA publicó el retiro Z-2273-2012 el 5 de septiembre de 2012. Peligro reportado: On May 9th 2012 Alphatec Spine received a complaint which reported that the distal tip of the Trestle Luxe Slide Alignment Tool (Part Number 71721), does not properly fit into the hexalobe screw on the Trestle Luxe Plate.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter