BioMerieux VIDAS System in combination with Quality Control VIDAS* (QCV*) Product Usage: the Quality Control VIDAS (QCV) is an automated test for use on the VIDAS® system to detect abnormal operation of the VIDAS and mini VIDAS instrument pipette mechanisms and optical systems. Specifically the mini VIDAS system is an immunodiagnostic system intended to be used by trained and qualified laboratory professionals, for In vitro Diagnostic (IVD) purpose, veterinary and industrial applications. The mini VIDAS® system is intended to execute an immunoassay protocol and to release results according to the package insert of the VIDAS® assay kits.

Publicado el 18 de julio de 2018 · Inicio del retiro: 4 de enero de 2018

Resumen

FDA registra el retiro Z-2429-2018 de dispositivos médicos de Clase II, publicado el 18 de julio de 2018 e iniciado el 4 de enero de 2018. Empresa responsable: bioMerieux, Inc.; Estado: Finalizado; Cantidad reportada: 1,866 units; Lote: Description, Reference No. (UDI No.): VIDAS ANALYSER 110/220V, 93296 (03573026133573); VIDAS ANALYSER 220V, 93297 (03573026133580); VIDAS ANALYSEUR, 99011 (03573026109257); VIDAS ANALYSER EU, 99444 (03573026114060); VIDAS ANALYSER US, 99445 (03573026114084); VIDAS ANALYSER, 99735 (03573026140427); VIDAS ANALYSER US, W2205 (03573026029678); VIDAS ANALYSER, 410417 (03573026351663); Quality Control VIDAS® (QCV), 30706 (03573026121631); VIDAS ANALYSER EU, W3205 (03573026030292); VIDAS INDUSTRY US, 421002 (03573026553319). La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Increase of events affecting performance regarding pump clogging and displacement of the color-coded SPR label (DOT). These performance issues could contribute to erroneous results for any assay performed on the VIDAS® and mini VIDAS® system due to a decrease of pipetted volume.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
bioMerieux, Inc.
Distribuidor
Worldwide Distribution - US Nationwide in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, VT, WA, WI, WV, WY. International distribution to Algeria, Angola, Anguilla, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bolivia, Bosnia - Herzegovina, Brazil, Brunei, Bulgaria, Burkina Faso, Cambodia, Cameroon, Canada, Cape Verde, Chad, Chile, China / Hong Kong, Colombia, Congo, Costa Rica, Croatia, Czech Republic, Denmark, Djibouti, Dominican Republic, Dutch Antilles, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Finland, France, Gabon, Georgia, Germany, Ghana, Greece, Guam, Guatemala, Guinea, Honduras, Hungary, India, Indonesia, Iran, Iraq, Israel, Italy, Ivory Coast, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Mali, Malta, Mauretania, Mexico, Mongolia, Morocco, Mozambique, Myanmar, Nepal, Netherlands, Nicaragua, Niger, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Senegal, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Tanzania, Thailand, Togo, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Vietnam, Zimbabwe.
Lote
Description, Reference No. (UDI No.): VIDAS ANALYSER 110/220V, 93296 (03573026133573); VIDAS ANALYSER 220V, 93297 (03573026133580); VIDAS ANALYSEUR, 99011 (03573026109257); VIDAS ANALYSER EU, 99444 (03573026114060); VIDAS ANALYSER US, 99445 (03573026114084); VIDAS ANALYSER, 99735 (03573026140427); VIDAS ANALYSER US, W2205 (03573026029678); VIDAS ANALYSER, 410417 (03573026351663); Quality Control VIDAS® (QCV), 30706 (03573026121631); VIDAS ANALYSER EU, W3205 (03573026030292); VIDAS INDUSTRY US, 421002 (03573026553319)
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿BioMerieux VIDAS System in combination with Quality Control VIDAS* (QCV*) Product Usage: the Quality Control VIDAS (QCV) is an automated test for use on the VIDAS® system to detect abnormal operation of the VIDAS and mini VIDAS instrument pipette mechanisms and optical systems. Specifically the mini VIDAS system is an immunodiagnostic system intended to be used by trained and qualified laboratory professionals, for In vitro Diagnostic (IVD) purpose, veterinary and industrial applications. The mini VIDAS® system is intended to execute an immunoassay protocol and to release results according to the package insert of the VIDAS® assay kits. fue retirado del mercado?

Sí. FDA publicó el retiro Z-2429-2018 el 18 de julio de 2018. Peligro reportado: Increase of events affecting performance regarding pump clogging and displacement of the color-coded SPR label (DOT). These performance issues could contribute to erroneous results for any assay performed on the VIDAS® and mini VIDAS® system due to a decrease of pipetted volume.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter