FDAFDA · Alimentos y medicamentos·N.º oficial Z-2453-2012
Hologic Fluoroscan Mini C-Arm Model with InSight- FD Software Version 5.0.1and 5.0.2 The device is a Mini C-arm Fluoroscopic Imaging System designed to provide -physicians with general fluoroscopic visualization of a patient, including but not limited to surgical orthopedic and podiatry use, critical and emergency procedures, and light anatomy imaging situation.
Publicado el 3 de octubre de 2012 · Inicio del retiro: 14 de septiembre de 2012
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Fluoroscan Mini C-Arm models -InSight2, version 4.0 and 4.0.1 - DICOM images stored in PACS may be flipped 180¿around vertical axis. InSight FD, version 5.0.1 and 5.0.2 - DICOM images stored in PACS may be lipped 180¿ around vertical axis as well as 180¿ around a horizontal axis.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Hologic, Inc.
- Distribuidor
- Worldwide Distribution - USA (nationwide) and the countries of Argentina, Australia,Belgium, Canada, China, Germany, Ecuador, Spain, France, Great Britain, Greece, Hong Kong, Italy, Japan Korea, Lithuvania, Neherlands, qatar, Russia, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Taiwan, South Africa, and Switzerland.
- Lote
- Software version 5.0.1and 5.0.2
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States