FDAFDA · Alimentos y medicamentos·N.º oficial Z-2468-2025
AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) SFCH CRANI PACK (PS 132432), MODEL/ITEM NUMBER UICR10W; 2) SFCH SHUNT PACK (PS 139798), MODEL/ITEM NUMBER UISH55V.
Publicado el 3 de septiembre de 2025 · Inicio del retiro: 14 de julio de 2025
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
Qué hacer
Voluntary: Firm initiated · E-Mail
Detalles del retiro
- Fabricante
- American Contract Systems Inc
- Distribuidor
- US distribution to states of: IL and LA
- Lote
- 1) MODEL/ITEM NUMBER UICR10W: UDI/DI 00191072207241, Kit Lot Numbers: 42-7828711, 42-7980711, 42-8063211, 42-8124011; 2) MODEL/ITEM NUMBER UISH55V: UDI/DI 00191072235442, Kit Lot Numbers: 42-7799111, 42-7985211, 42-8065111, 42-8140711.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States