Stryker Orthopaedics Restoration Modular Cone Body Trials Stryker Ireland Carrigtwohill Industrial Estate Carrigtwohill County Cork, Ireland. The Cone Body Trials simulate the respective Cone Body implants and mate with the stem trials to assist the surgeon in assessing the fit of the proposed Implants. The cone body trials is an assembly of the Body and 8 mm Hex Locking Bolt. The trial stem is attached to the trial Cone body via the distal thread of the Locking bolt.
Publicado el 10 de octubre de 2012 · Inicio del retiro: 2 de abril de 2012
Resumen
FDA registra el retiro Z-2487-2012 de dispositivos médicos de Clase II, publicado el 10 de octubre de 2012 e iniciado el 2 de abril de 2012. Empresa responsable: Stryker Howmedica Osteonics Corp.; Estado: Finalizado; Cantidad reportada: 10, 983 units; Lote: Catalog No. 6278-1-019 Mod Rev 19 mm std cone trial 6278-1-021 Mod Rev 21 mm std cone trial 6278-1-023 Mod Rev 23 mm std cone trial 6278-1-025 Mod Rev 25 mm std cone trial 6278-1-027 Mod Rev 27 mm std cone trial 6278-1-029 Mod Rev 29 mm std cone trial 6278-1-031 Mod Rev 31 mm std cone trial 6278-1-119 Mod Rev 19 mm + 10 mm cone trial 6278-1-121 Mod Rev 21mm + 10 mm cone trial 6278-1-123 Mod Rev 23mm + 10 mm cone trial 6278-1 -125 Mod Rev 25 mm + 10 mm cone trial 6278-1-127 Mod Rev 27mm + 10 mm cone trial 6278-1-129 Mod Rev 29mm + 10mm cone trial 6278-1-131 Mod Rev 31mm + 10 mm cone trial 6278 -1-219 Mod Rev 19mm + 20 mm cone trial 6278-1-221 Mod Rev 21mm +20 mm cone trial 6278-1-223 Mod Rev 23mm + 20 mm cone trial 6278-1-225 Mod Rev 25mm + 20 mm cone trial 6278-1-227 Mod Rev 27mm + 20 mm cone trial 6278-1-229 Mod Rev 29mm + 20 mm cone trial 6278-1-231 Mod Rev 31mm + 20 mm cone trial 6278-1-319 Mod Rev 19mm + 30 mm cone trial 6278-1-321 Mod Rev 21mm +30 mm cone trial 6278-1-323 Mod Rev 23mm + 30 mm cone trial 6278-1-325 Mod Rev 25mm + 30 mm cone trial 6278-1-327 Mod Rev 27mm + 30 mm cone trial 6278-1-329 Mod rev 29mm +30 mm cone trial 6278-1-331 Mod Rev 31mm + 30 mm cone trial 1 single unit to each non-sterile package Numerous lots numbers please see attachment. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Stryker Orthopaedics has become aware of a burr, or a metal shaving, observed in the internal threaded portion of the Cone Body Trial. The typical size of the shaving observed was 1-2 mm in length and 0.037 mm squared in cross section.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Stryker Howmedica Osteonics Corp.
- Distribuidor
- Nationwide distribution
- Lote
- Catalog No. 6278-1-019 Mod Rev 19 mm std cone trial 6278-1-021 Mod Rev 21 mm std cone trial 6278-1-023 Mod Rev 23 mm std cone trial 6278-1-025 Mod Rev 25 mm std cone trial 6278-1-027 Mod Rev 27 mm std cone trial 6278-1-029 Mod Rev 29 mm std cone trial 6278-1-031 Mod Rev 31 mm std cone trial 6278-1-119 Mod Rev 19 mm + 10 mm cone trial 6278-1-121 Mod Rev 21mm + 10 mm cone trial 6278-1-123 Mod Rev 23mm + 10 mm cone trial 6278-1 -125 Mod Rev 25 mm + 10 mm cone trial 6278-1-127 Mod Rev 27mm + 10 mm cone trial 6278-1-129 Mod Rev 29mm + 10mm cone trial 6278-1-131 Mod Rev 31mm + 10 mm cone trial 6278 -1-219 Mod Rev 19mm + 20 mm cone trial 6278-1-221 Mod Rev 21mm +20 mm cone trial 6278-1-223 Mod Rev 23mm + 20 mm cone trial 6278-1-225 Mod Rev 25mm + 20 mm cone trial 6278-1-227 Mod Rev 27mm + 20 mm cone trial 6278-1-229 Mod Rev 29mm + 20 mm cone trial 6278-1-231 Mod Rev 31mm + 20 mm cone trial 6278-1-319 Mod Rev 19mm + 30 mm cone trial 6278-1-321 Mod Rev 21mm +30 mm cone trial 6278-1-323 Mod Rev 23mm + 30 mm cone trial 6278-1-325 Mod Rev 25mm + 30 mm cone trial 6278-1-327 Mod Rev 27mm + 30 mm cone trial 6278-1-329 Mod rev 29mm +30 mm cone trial 6278-1-331 Mod Rev 31mm + 30 mm cone trial 1 single unit to each non-sterile package Numerous lots numbers please see attachment.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAHoffmann LRF Safety Clip (Bone Transport Strut), 14 x 12 x 5 mm, REF 4933-0-401, NDC31 de mayo de 2017
- FDASupracondylar Nail T2 SCN 09x300 mm T2 Supracondylar Nail System is intended to provide strong and stable21 de septiembre de 2016
- FDAStryker Orthopaedics Specialty Triathlon Tibial Alignment Handle with Secondary Lock, Non-Sterile, Used to17 de agosto de 2016
- FDAStryker, T2 Guidewire, Ball-Tipped, 3 x 1000mm Product Usage: Intended for the temporary stabilization of bone2 de marzo de 2016
- FDAOptilock T15 AO Driver (3.5mm). Intended Use: The OptiLock Upper Extremity Plating System is intended for10 de octubre de 2012
- FDATabs Professional Monitor, includes one wall bracket, Model 25022. Stanley Security Solutions, Inc., Lincoln,10 de octubre de 2012
Preguntas sobre este retiro
¿Stryker Orthopaedics Restoration Modular Cone Body Trials Stryker Ireland Carrigtwohill Industrial Estate Carrigtwohill County Cork, Ireland. The Cone Body Trials simulate the respective Cone Body implants and mate with the stem trials to assist the surgeon in assessing the fit of the proposed Implants. The cone body trials is an assembly of the Body and 8 mm Hex Locking Bolt. The trial stem is attached to the trial Cone body via the distal thread of the Locking bolt. fue retirado del mercado?
Sí. FDA publicó el retiro Z-2487-2012 el 10 de octubre de 2012. Peligro reportado: Stryker Orthopaedics has become aware of a burr, or a metal shaving, observed in the internal threaded portion of the Cone Body Trial. The typical size of the shaving observed was 1-2 mm in length and 0.037 mm squared in cross section.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter