Leica HistoCore Arcadia H, REF: 14039354100 and 14039357258, Paraffin Embedding Station.
Publicado el 14 de agosto de 2024 · Inicio del retiro: 10 de mayo de 2024
Resumen
FDA registra el retiro Z-2529-2024 de dispositivos médicos de Clase II, publicado el 14 de agosto de 2024 e iniciado el 10 de mayo de 2024. Empresa responsable: LEICA BIOSYSTEMS NUSSLOCH GMBH; Estado: En curso; Cantidad reportada: 180 devices; Lote: Model/Catalog Number: 14039354100; UDI-DI: 04049188194763, Serial numbers: 58, 62, 98, 99, 101, 117,120, 121, 122, 123, 124, 125, 127, 128, 129, 177, 178, 179, 180, 181, 185, 186, 218, 222, 223, 279, 282, 284, 286, 287, 288, 289, 290, 291, 292, 314, 315, 317, 318, 319, 321, 322, 353, 354, 355, 356, 357, 358, 360, 363, 365, 366, 371, 478, 480, 483, 484, 487, 488, 489, 490, 491, 492, 553, 560, 561, 562, 584, 586, 588, 589, 590, 593, 597, 643, 644, 647, 649, 650, 651, 652, 653, 656, 658, 659, 697, 698, 705, 710, 714, 715, 716, 717, 739, 740, 741, 742, 743, 757, 819, 820, 821, 822, 825, 826, 827, 828, 829, 830, 831, 832, 878, 879, 880, 881, 882, 883, 884, 885, 886, 887, 888, 889, 890, 891, 892, 893, 894, 896, 897, 898, 900, 901, 904, 906, 907, 909, 911, 919, 921, 922, 923, 924, 927, 928, 930, 1113, 1114, 1115, 1116, 1117, 1118, 1123, 1124, 1198, 1199, 1200, 1201, 1202, 1203, 1204, 1205, 1206, 1207, 1212, 1216, 1217, 1266, 1267, 1268, 1271, 1273, 1274, 1277, 1278, 1279, 1280, 1283, and 1285. Model/Catalog Number: 14039357258; UDI-DI: 04049188205971, serial number 5033. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
An issue with safe usage of the device was identified whereby toxic smoke and internal fire developed resulting from a problem associated with the sealing of the paraffin tank.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- LEICA BIOSYSTEMS NUSSLOCH GMBH
- Distribuidor
- US Nationwide, including government distribution.
- Lote
- Model/Catalog Number: 14039354100; UDI-DI: 04049188194763, Serial numbers: 58, 62, 98, 99, 101, 117,120, 121, 122, 123, 124, 125, 127, 128, 129, 177, 178, 179, 180, 181, 185, 186, 218, 222, 223, 279, 282, 284, 286, 287, 288, 289, 290, 291, 292, 314, 315, 317, 318, 319, 321, 322, 353, 354, 355, 356, 357, 358, 360, 363, 365, 366, 371, 478, 480, 483, 484, 487, 488, 489, 490, 491, 492, 553, 560, 561, 562, 584, 586, 588, 589, 590, 593, 597, 643, 644, 647, 649, 650, 651, 652, 653, 656, 658, 659, 697, 698, 705, 710, 714, 715, 716, 717, 739, 740, 741, 742, 743, 757, 819, 820, 821, 822, 825, 826, 827, 828, 829, 830, 831, 832, 878, 879, 880, 881, 882, 883, 884, 885, 886, 887, 888, 889, 890, 891, 892, 893, 894, 896, 897, 898, 900, 901, 904, 906, 907, 909, 911, 919, 921, 922, 923, 924, 927, 928, 930, 1113, 1114, 1115, 1116, 1117, 1118, 1123, 1124, 1198, 1199, 1200, 1201, 1202, 1203, 1204, 1205, 1206, 1207, 1212, 1216, 1217, 1266, 1267, 1268, 1271, 1273, 1274, 1277, 1278, 1279, 1280, 1283, and 1285. Model/Catalog Number: 14039357258; UDI-DI: 04049188205971, serial number 5033.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- Germany
Retiros relacionados
- FDALeica Biosystems Cryostat Model Number CM1100, Material Number 14046931130, for freezing and sectioning tissue23 de octubre de 2024
- FDALeica Biosystems Cryostat Model Number CM1950, Material Numbers: a) 14047742467, b) 1491950C1US, c)23 de octubre de 2024
- FDALeica Biosystems Cryostat Model Number CM1900, Material Numbers: a) 14045227382, b) 14901900100, c)23 de octubre de 2024
- FDALeica Biosystems Cryostat Model Number CM1850, Material Numbers: a) 14047131148, b) 14901850001, c)23 de octubre de 2024
- FDAMultitom Rax -Intended to visualize anatomical structures by converting an x-ray pattern into a visible image14 de agosto de 2024
- FDABaxter SIGMA Spectrum Infusion Pump, Product Code: 35700BAX214 de agosto de 2024
Preguntas sobre este retiro
¿Leica HistoCore Arcadia H, REF: 14039354100 and 14039357258, Paraffin Embedding Station. fue retirado del mercado?
Sí. FDA publicó el retiro Z-2529-2024 el 14 de agosto de 2024. Peligro reportado: An issue with safe usage of the device was identified whereby toxic smoke and internal fire developed resulting from a problem associated with the sealing of the paraffin tank.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter