Monaco RTP System, 5.50/5.51, radiation treatment planning system Product Usage: The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy. The system calculates dose for photon, electron, and carbon treatment plans and displays, on-screen and in hard-copy, two- or three-dimensional radiation dose distributions inside patients for given treatment plan set-ups. The Monaco product line is intended for use in radiation treatment planning. It uses generally accepted methods for: -contouring -image manipulation -simulation -image fusion -plan optimization -QA and plan review
Publicado el 2 de octubre de 2019 · Inicio del retiro: 4 de septiembre de 2019
Resumen
FDA registra el retiro Z-2561-2019 de dispositivos médicos de Clase II, publicado el 2 de octubre de 2019 e iniciado el 4 de septiembre de 2019. Empresa responsable: Elekta Inc; Estado: En curso; Cantidad reportada: 49 units; Lote: UDI: (01)00858164002244(10) 5.50.00 (01)00858164002268(10) 5.51.00. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
It is possible that the forced electron density settings will be changed for some structures unintentionally and this can result in incorrect dose calculation.
Qué hacer
Voluntary: Firm initiated · E-Mail
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Elekta Inc
- Distribuidor
- Worldwide Distribution - US Nationwide in the states of WI, OH, NY, OH, IN, IA, TX, and countries of Canada, China, Denmark, Germany, Greece, Hong Kong, Italy, Netherlands, Sweden, UK
- Lote
- UDI: (01)00858164002244(10) 5.50.00 (01)00858164002268(10) 5.51.00
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAC-ALBUMIN 5% OPHTH SOLUTION, 10 mL droptainer bottles, Rx only, Colonia Care Pharmacy, 515 Inman Ave., Suite9 de octubre de 2019
- FDAC-ATROPINE SULF OPHTH 0.02%, 5 mL droptainer bottles, Rx only, Colonia Care Pharmacy, 515 Inman Ave., Suite A,9 de octubre de 2019
- FDARandox Urinalysis Control Level 2 (IVD) Ref: UC5034 - Product Usage: Product is intended for in vitro2 de octubre de 2019
- FDAAccu-Chek Connect Diabetes Management App2 de octubre de 2019
- FDAMULTIGENT Lithium-intended for the quantitation of lithium in serum or plasma using respectively the ARCHITECT2 de octubre de 2019
- FDADecorated Red Apple Cookies sold in display case in retail stores2 de octubre de 2019
Preguntas sobre este retiro
¿Monaco RTP System, 5.50/5.51, radiation treatment planning system Product Usage: The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy. The system calculates dose for photon, electron, and carbon treatment plans and displays, on-screen and in hard-copy, two- or three-dimensional radiation dose distributions inside patients for given treatment plan set-ups. The Monaco product line is intended for use in radiation treatment planning. It uses generally accepted methods for: -contouring -image manipulation -simulation -image fusion -plan optimization -QA and plan review fue retirado del mercado?
Sí. FDA publicó el retiro Z-2561-2019 el 2 de octubre de 2019. Peligro reportado: It is possible that the forced electron density settings will be changed for some structures unintentionally and this can result in incorrect dose calculation.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · E-Mail