Brilliance BigBore Radiology CT Model number 728244 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.

Publicado el 31 de agosto de 2016 · Inicio del retiro: 20 de julio de 2016

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at iso-center (0,0). If the region of interest is not visualized in the images, a CT rescan may be performed.

Qué hacer

Voluntary: Firm initiated · E-Mail

Detalles del retiro

Fabricante
Philips Medical Systems (Cleveland) Inc
Distribuidor
US Nationwide Distribution in the states of: AL,AR,AZ,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA, MD,ME,MI,MN,MO,MS,NC,ND,NE,NJ,NM, NV,NY,OH,OK,OR,PA,PR,SC,TN,TX,UT,VA,VT,WA,WI,WV,WY
Lote
728244  Brilliance BigBore Radiology CT 7375 7781 7614 7633 7708 7789 7721 7760 7252 7380 7390 7514 7052 7244 7384 7533 7172 7200 7546 7427 7452 7407 7453 7586
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States