Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-2693-2014

Minstrel (with scale) Product Usage: is a mobile passive hoist. It is intended for lifting and transporting adult residents on horizontal floors in hospitals, nursing homes or other health care facilitates. The Minstrel device aids in the lifting, transfer and lowering to safe position, of a patient. This is accomplished by placing a sling under the patient, and attaching the sling to a spreader bar that is connected to a lifting arm jig via a connection to the spreader bar that allows a swiveling up-down movement. Using battery power, an actuator then powers the lifting arm upwards so that the patient in the sling is lifted from, typically, a bed or chair.

Publicado el 1 de octubre de 2014 · Inicio del retiro: 28 de julio de 2014

Resumen

FDA registra el retiro Z-2693-2014 de dispositivos médicos de Clase II, publicado el 1 de octubre de 2014 e iniciado el 28 de julio de 2014. Empresa responsable: ARJOHUNTLEIGH POLSKA Sp. z.o.o.; Estado: Finalizado; Cantidad reportada: 18 Lifts; Lote: Model Numbers: HMA0002 and HMB002-US; Serial Numbers: MPLAT0409W1177 ; MPLAT0410W1213 ; MPLAT0410W1228 ; MPLAT0410W1229 ; MPLAT0410W1231 ; MPLAT0410W1232 ; MPLAT0412W1290 ; MPLAT0412W1296 ; MPLAT0412W1297 ; MPLAT0502W1403 ; MPLAT0502W1404 ; MPLAT0503W1485 ; MPLAT0503W1486 ; MPLAT0504W1518 ; MPLAT0504W1519 ; MPLAT0504W1520 ; MPLAT0504W1531. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

ArjoHuntleigh received three reports where the bolt under the scale unscrewed from the scale. There is a remote risk concerning the bolt unscrewing from the scale. When the scale and bolt parts are not correctly serviced, or when they receive a sharp blow during or outside of use, it is possible the bolt holding the spreader bar to the scale could unscrew during subsequent use. As a result, the sp

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
ARJOHUNTLEIGH POLSKA Sp. z.o.o.
Distribuidor
Worldwide Distribution - US Nationwide in the state of MA and the countries of Australia, Germany, and Italy.
Lote
Model Numbers: HMA0002 and HMB002-US; Serial Numbers: MPLAT0409W1177 ; MPLAT0410W1213 ; MPLAT0410W1228 ; MPLAT0410W1229 ; MPLAT0410W1231 ; MPLAT0410W1232 ; MPLAT0412W1290 ; MPLAT0412W1296 ; MPLAT0412W1297 ; MPLAT0502W1403 ; MPLAT0502W1404 ; MPLAT0503W1485 ; MPLAT0503W1486 ; MPLAT0504W1518 ; MPLAT0504W1519 ; MPLAT0504W1520 ; MPLAT0504W1531
Clasificación de riesgo
Class II
Área (FDA)
device
País
Poland

Retiros relacionados

Preguntas sobre este retiro

¿Minstrel (with scale) Product Usage: is a mobile passive hoist. It is intended for lifting and transporting adult residents on horizontal floors in hospitals, nursing homes or other health care facilitates. The Minstrel device aids in the lifting, transfer and lowering to safe position, of a patient. This is accomplished by placing a sling under the patient, and attaching the sling to a spreader bar that is connected to a lifting arm jig via a connection to the spreader bar that allows a swiveling up-down movement. Using battery power, an actuator then powers the lifting arm upwards so that the patient in the sling is lifted from, typically, a bed or chair. fue retirado del mercado?

Sí. FDA publicó el retiro Z-2693-2014 el 1 de octubre de 2014. Peligro reportado: ArjoHuntleigh received three reports where the bolt under the scale unscrewed from the scale. There is a remote risk concerning the bolt unscrewing from the scale. When the scale and bolt parts are not correctly serviced, or when they receive a sharp blow during or outside of use, it is possible the bolt holding the spreader bar to the scale could unscrew during subsequent use. As a result, the sp

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter