Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-2817-2018

bioMerieux VITEK 2 Gram Positive Cefoxitin Screen, VITEK 2 Gram Positive AST for Oxacillin, VITEK SYSTEMS General Susceptibility Card Cefotoxin and AST for Oxacillin bioMerieux VITEK 2 AST-P631 bioMerieux VITEK 2 AST-GP67 Test Kit bioMerieux VITEK 2 AST-GP71 Test Kit bioMerieux VITEK 2 AST-GP75 Test Kit bioMerieux VITEK 2 AST-GP78 Test Kit Components for an automated invitro diagnostic system consisting of instruments, software and reagent cards designed for the ID and antimicrobial susceptibility testing of bacteria and yeast.

Publicado el 29 de agosto de 2018 · Inicio del retiro: 26 de enero de 2018

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Customer reports indicated an increase in the rate of non-detected MRSA in association with the VITEK 2 AST-P631 product.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
bioMerieux, Inc.
Distribuidor
Worldwide distribution. US nationwide including Puerto Rico, Bermuda, Brazil, Canada, China, Curacao, Costa Rica, Dominican Republic, Ecuador, France, Guatemala, Indonesia, Israel, Hong Kong, South Korea, South Africa, Mexico, Myanmar, Nicaragua, Peru, Philippines, Paraguay, Singapore, Thailand, El Salvador, and Vietnam.
Lote
VITEK 2 AST-P631, Ref # 414961 UDI 03573026426804 OXSF, multiple Gram Positive Cards all lots of cards with Oxacillin and Cefotoxin OX, OXSF, multiple Gram Positive Cards all lots of cards with Oxacillin and Cefotoxin VITEK 2 AST-GP67 Test Kit, Ref: # 22226 UDI 03573026242060, OX, OXSF, multiple Gram Positive Cards, all lots of cards with Oxacillin and Cefotoxin. VITEK 2 AST-GP71 Test Kit, Ref: #410750 UDI 03573026355173, OX, OXSF, multiple Gram Positive Cards, all lots of cards with Oxacillin and Cefotoxin. VITEK 2 AST-GP75 Test Kit, Ref: #415670 UDI 03573026437183, OX, OXSF, multiple Gram Positive Cards, all lots of cards with Oxacillin and Cefotoxin. VITEK 2 AST-GP78 Test Kit, Ref: #421051 UDI 03573026553982, OX, OXSF, multiple Gram Positive Cards, all lots of cards with Oxacillin and Cefotoxin.
Clasificación de riesgo
Class I
Área (FDA)
device
País
United States

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta