Positron Emission Tomography and Computed Tomography System, Model: uMI 550, Material Number: 88000057 - Product Usage: This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.

Publicado el 26 de agosto de 2020 · Inicio del retiro: 26 de marzo de 2020

Resumen

FDA registra el retiro Z-2819-2020 de dispositivos médicos de Clase II, publicado el 26 de agosto de 2020 e iniciado el 26 de marzo de 2020. Empresa responsable: Shanghai United Imaging Healthcare Co., Ltd.; Estado: Finalizado; Cantidad reportada: 3; Lote: UDI: 06971576832026, Software Version: R001.3.0.0.750505, Serial Numbers: 200017, 200023, 200024. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

In affected software version, of Positron emission tomography and computed tomography system, the CT Hounsfield unit curve may not display in real time when using the bolus tracking protocol with the tracker scan function. If this occurs, the subsequent clinical scanning protocols will need to be started manually, which may effect image enhancement, and may result in patient rescan.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Shanghai United Imaging Healthcare Co., Ltd.
Distribuidor
US Nationwide distribution including in the state of TX.
Lote
UDI: 06971576832026, Software Version: R001.3.0.0.750505, Serial Numbers: 200017, 200023, 200024
Clasificación de riesgo
Class II
Área (FDA)
device
País
China

Retiros relacionados

Preguntas sobre este retiro

¿Positron Emission Tomography and Computed Tomography System, Model: uMI 550, Material Number: 88000057 - Product Usage: This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology. fue retirado del mercado?

Sí. FDA publicó el retiro Z-2819-2020 el 26 de agosto de 2020. Peligro reportado: In affected software version, of Positron emission tomography and computed tomography system, the CT Hounsfield unit curve may not display in real time when using the bolus tracking protocol with the tracker scan function. If this occurs, the subsequent clinical scanning protocols will need to be started manually, which may effect image enhancement, and may result in patient rescan.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter