Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-2962-2024

Impella 5.5 with SmartAssist Set, CA; Product Code: 0550-0004;

Publicado el 11 de septiembre de 2024 · Inicio del retiro: 5 de agosto de 2024

Resumen

FDA registra el retiro Z-2962-2024 de dispositivos médicos de Clase II, publicado el 11 de septiembre de 2024 e iniciado el 5 de agosto de 2024. Empresa responsable: Abiomed, Inc.; Estado: En curso; Cantidad reportada: 21 units; Lote: Product Code: 0550-0004; UDI-DI: 00813502010466; Serial Numbers: 439898 437211A 437204A 440938 439938 439939 453784 453785 476355 476356 470678 493496 493497 491749 491752 491751 478569A 483483A 491750 485277A 485281A; Batch Numbers: 2024337828 2024338391 2024338394 2024340273 2024341579 2024341586 2024356969 2024356971 2024397878 2024397886 2024399646 2025440057 2025440061 2025448005 2025452639 2025452668 2025484262 2025484276 2025448006 2025500661 2025500668. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Abiomed, Inc.
Distribuidor
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Brunei Darussalam, Canada, Croatia, Czeck Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Panama, Poland, Saudi Arabia, Serbia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, U.A.E., United Kingdom.
Lote
Product Code: 0550-0004; UDI-DI: 00813502010466; Serial Numbers: 439898 437211A 437204A 440938 439938 439939 453784 453785 476355 476356 470678 493496 493497 491749 491752 491751 478569A 483483A 491750 485277A 485281A; Batch Numbers: 2024337828 2024338391 2024338394 2024340273 2024341579 2024341586 2024356969 2024356971 2024397878 2024397886 2024399646 2025440057 2025440061 2025448005 2025452639 2025452668 2025484262 2025484276 2025448006 2025500661 2025500668;
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Impella 5.5 with SmartAssist Set, CA; Product Code: 0550-0004; fue retirado del mercado?

Sí. FDA publicó el retiro Z-2962-2024 el 11 de septiembre de 2024. Peligro reportado: A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter