FDAFDA · Alimentos y medicamentos·N.º oficial Z-3051-2020
Drug Screening Component: Uncut Sheets Part Number/ Description/ LOT Number 100163/ EDI OPI (300) Uncut Sheet-Card /187876, 100195/ EDI PPX Uncut Sheet-Card/ 187877, 160130/ EDI COC Uncut Sheet-Cup/ 188168, 160180/ EDI PCP Uncut Sheet-Cup/ 189027, 160140/ EDI THC Uncut Sheet Cup /189028, 160170/ EDI BZO Uncut Sheet-Cup/ 189081, 160120/ EDI MET1000 Uncut Sheet-Cup/ 189082, 160160/ EDI OPI2000 Uncut Sheet-Cup / 189123, 160130/ EDI COC Uncut Sheet-Cup/ 189124, 160160/ EDI OPI2000 Uncut Sheet-Cup/ 189166, 160163/ EDI OPI (300) Uncut Sheet-Cup/ 189191, 160110/ EDI AMP Uncut Sheet-Cup/ 189371, 160140/ EDI THC Uncut Sheet Cup/ 189372, 160123/ EDI MET(300) Uncut Sheet-Cup/ 189425, 160170/ EDI BZO Uncut Sheet-Cup/ 189426, 100163/ EDI OPI (300) Uncut Sheet-Card/ 189456, 160170/ EDI BZO Uncut Sheet-Cup/ 189565, 160180/ EDI PCP Uncut Sheet-Cup/ 189601, 160125/ EDI MET(500) Uncut Sheet-Cup/ 189748, 160130/ EDI COC Uncut Sheet-Cup/ 189786, 160160/ EDI OPI2000 Uncut Sheet-Cup / 189844, 160180/ EDI PCP Uncut Sheet-Cup/ 189937, 160172/ EDI BZO(200) Uncut Sheet-CUP/189956 Updated 09/03/2020: EDI Affected products -Catalog Numbers/ Lot Numbers 31105-3/ PI-01081 60503s-3ZC/ PH-10132 60600/ PH-11005
Publicado el 7 de octubre de 2020 · Inicio del retiro: 18 de agosto de 2020
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Ameditech Inc
- Distribuidor
- US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
- Lote
- ALL Lots Within Expiry
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States