Merge Unity software, formerly known as DR Systems Unity PACS software. The firm name on the label is Merge Healthcare, Hartland, WI. Merge Unity is a medical image and information management system that allows viewing, selection, processing, printing, telecommunications, and media interchange of medical images from a variety of diagnostic imaging systems
Publicado el 13 de septiembre de 2017 · Inicio del retiro: 14 de agosto de 2017
Resumen
FDA registra el retiro Z-3105-2017 de dispositivos médicos de Clase II, publicado el 13 de septiembre de 2017 e iniciado el 14 de agosto de 2017. Empresa responsable: Merge Healthcare, Inc.; Estado: Finalizado; Cantidad reportada: 98 sites have the potentially affected software; Lote: Software versions 10.0, 10.0.1, 10.0.2, 10.0.3, 10.0.4, 10.0.5, 11.0, 11.0.1, 11.0.2, 11.0.3, 11.0.4, 11.1, 11.1.1, 11.1.1.1, 11.1.1 Patch 2, 11.1.1 Patch 3, 11.1.2, 11.1.2 Patch 1, 11.1.2 Patch 2, 11.1.2 Patch 3, and 11.1.2 Patch 4. Only customers with one of the above listed software versions AND a license for Mammo Tracking are impacted by this recall. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The software is not identifying the patient as having atypical hyperplasia, resulting in an incorrect Gail Risk calculation.
Qué hacer
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Merge Healthcare, Inc.
- Distribuidor
- Nationwide Distribution
- Lote
- Software versions 10.0, 10.0.1, 10.0.2, 10.0.3, 10.0.4, 10.0.5, 11.0, 11.0.1, 11.0.2, 11.0.3, 11.0.4, 11.1, 11.1.1, 11.1.1.1, 11.1.1 Patch 2, 11.1.1 Patch 3, 11.1.2, 11.1.2 Patch 1, 11.1.2 Patch 2, 11.1.2 Patch 3, and 11.1.2 Patch 4. Only customers with one of the above listed software versions AND a license for Mammo Tracking are impacted by this recall.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAMerge PACS and Merge OrthoPACS, medical imaging diagnostic workstation2 de septiembre de 2026
- FDAMerge OrthoPACS, medical imaging diagnostic workstation12 de agosto de 2026
- FDAMerge PACS, medical imaging diagnostic workstation12 de agosto de 2026
- FDAVERICIS, Merge Cardio, Model/Catalog Number: Version 9.x, Software Version: 9.0.6 and 9.0.8; VERICIS is an26 de marzo de 2025
- FDAJTS External Drive Unit JTS Distal Femoral Extendible Implant Femur is intended for the replacement of13 de septiembre de 2017
- FDAPresource(R) Kits (custom surgical/procedural kits): The standard and custom kits are packed with all of the13 de septiembre de 2017
Preguntas sobre este retiro
¿Merge Unity software, formerly known as DR Systems Unity PACS software. The firm name on the label is Merge Healthcare, Hartland, WI. Merge Unity is a medical image and information management system that allows viewing, selection, processing, printing, telecommunications, and media interchange of medical images from a variety of diagnostic imaging systems fue retirado del mercado?
Sí. FDA publicó el retiro Z-3105-2017 el 13 de septiembre de 2017. Peligro reportado: The software is not identifying the patient as having atypical hyperplasia, resulting in an incorrect Gail Risk calculation.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit