FDAFDA · Alimentos y medicamentos·N.º oficial Z-3134-2017

VITEK(R) 2 GP, REF 21342, IVD, UDI 03573026131920, bioMerieux, Inc. VITEK(R) 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK¿ 2 utilizes growth based biochemical patterns to determine identification. The VITEK¿ 2 provides minimal inhibitory concentration (MIC) results for most organism/drug combinations as well as a category interpretation (S, I, or R) consistent with accepted criteria

Publicado el 27 de septiembre de 2017 · Inicio del retiro: 25 de agosto de 2017

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Customer reports have indicated occurrences of atypical negative O129R reactions for Enterococcus casseliflavus ATCC(R) 700327(TM) in association with VITEK(R) 2 GP ID Lot 2420240403 and 2420303403, causing failure of the Quality Control (QC) testing for the VITEK(R) 2 GP ID card. The expected O129R reaction for ATCC¿ 700327(TM) strain is positive.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Biomerieux Inc
Distribuidor
US and Angola, Argentina, Austria, Belarus, Belgium, Burkina Faso, Cameroon, Canada, China, Congo, Croatia, Egypt, France, Germany, Greece, Guam, Hungary, Israel, Italy, Latvia, Luxembourg, Macedonia, Mali, Mexico, Morocco, Netherlands, Nicaragua, Oman, Panama, Poland, Saudi Arabia, Serbia, South Africa, South Korea, Spain, Swaziland, Switzerland, Tunisia, Turkey, Turkmenistan, United Arab Emirates, United Kingdom
Lote
Lot Number 2420240403, Exp. 24-JUL-18, 2420303403, Exp 24-SEP-2018
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States