VITEK(R) 2 GP, REF 21342, IVD, UDI 03573026131920, bioMerieux, Inc. VITEK(R) 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK® 2 utilizes growth based biochemical patterns to determine identification. The VITEK® 2 provides minimal inhibitory concentration (MIC) results for most organism/drug combinations as well as a category interpretation (S, I, or R) consistent with accepted criteria
Publicado el 27 de septiembre de 2017 · Inicio del retiro: 25 de agosto de 2017
Resumen
FDA registra el retiro Z-3134-2017 de dispositivos médicos de Clase II, publicado el 27 de septiembre de 2017 e iniciado el 25 de agosto de 2017. Empresa responsable: Biomerieux Inc; Estado: Finalizado; Cantidad reportada: 11,703 cartons; Lote: Lot Number 2420240403, Exp. 24-JUL-18, 2420303403, Exp 24-SEP-2018. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Customer reports have indicated occurrences of atypical negative O129R reactions for Enterococcus casseliflavus ATCC(R) 700327(TM) in association with VITEK(R) 2 GP ID Lot 2420240403 and 2420303403, causing failure of the Quality Control (QC) testing for the VITEK(R) 2 GP ID card. The expected O129R reaction for ATCC® 700327(TM) strain is positive.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Biomerieux Inc
- Distribuidor
- US and Angola, Argentina, Austria, Belarus, Belgium, Burkina Faso, Cameroon, Canada, China, Congo, Croatia, Egypt, France, Germany, Greece, Guam, Hungary, Israel, Italy, Latvia, Luxembourg, Macedonia, Mali, Mexico, Morocco, Netherlands, Nicaragua, Oman, Panama, Poland, Saudi Arabia, Serbia, South Africa, South Korea, Spain, Swaziland, Switzerland, Tunisia, Turkey, Turkmenistan, United Arab Emirates, United Kingdom
- Lote
- Lot Number 2420240403, Exp. 24-JUL-18, 2420303403, Exp 24-SEP-2018
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAVITEK 2 AST-N461 TEST KIT, GRAM NEGATIVE SUSCEPTIBILITY TESTING - AUTOMATED, Reference Number 42481130 de septiembre de 2026
- FDAVITEK 2 AST-N447 TEST KIT Product reference 42462015 de julio de 2026
- FDAVITEK 2 Gram-negative Susceptibility Cards Containing o Polymyxin B (pb02n) o Ceftazidime/avibactam (cza02n) o11 de febrero de 2026
- FDAVITEK 2 - AST Test kits (in vitro diagnostic test cards for antimicrobial susceptibility): a) AST-N334, REF1 de mayo de 2024
- FDALaxmi Rusks, 10.58 OZ (300 gram) bags; UPC Code: 7 23246 26482 027 de septiembre de 2017
- FDAAQUIOS CL Flow Cytometer System The AQUIOS CL Flow Cytometer is intended for use as an in vitro diagnostic27 de septiembre de 2017
Preguntas sobre este retiro
¿VITEK(R) 2 GP, REF 21342, IVD, UDI 03573026131920, bioMerieux, Inc. VITEK(R) 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK® 2 utilizes growth based biochemical patterns to determine identification. The VITEK® 2 provides minimal inhibitory concentration (MIC) results for most organism/drug combinations as well as a category interpretation (S, I, or R) consistent with accepted criteria fue retirado del mercado?
Sí. FDA publicó el retiro Z-3134-2017 el 27 de septiembre de 2017. Peligro reportado: Customer reports have indicated occurrences of atypical negative O129R reactions for Enterococcus casseliflavus ATCC(R) 700327(TM) in association with VITEK(R) 2 GP ID Lot 2420240403 and 2420303403, causing failure of the Quality Control (QC) testing for the VITEK(R) 2 GP ID card. The expected O129R reaction for ATCC® 700327(TM) strain is positive.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter