FDAFDA · Alimentos y medicamentos·N.º oficial Z-3194-2024

Ingenia 1.5T -For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 1) 782115 2) 782101 3) 782140

Publicado el 2 de octubre de 2024 · Inicio del retiro: 3 de septiembre de 2024

Resumen

FDA registra el retiro Z-3194-2024 de dispositivos médicos de Clase II, publicado el 2 de octubre de 2024 e iniciado el 3 de septiembre de 2024. Empresa responsable: Philips North America Llc; Estado: En curso; Cantidad reportada: 48 units; Lote: Model UDI-DI: 782115 (01)00884838099043(21); 782101 (01)00884838098275(21); 782140 (01)00884838108646(21) Serial Numbers: 45601 61291 61312 61315 61328 61336 61337 61339 61348 61349 61351 61354 61355 61356 61357 61359 61360 61361 61362 61363 61364 61367 61371 61373 61374 61375 61376 61377 61378 61380 61383 61384 61394 61395 61397 61403 61404 61406 61407 61412 61415 61416 87645 87648 87649 87651 87654 87655. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Philips North America Llc
Distribuidor
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Iceland, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Maldives, Mauritius, Mexico, Morocco, Netherlands, Pakistan, Peru, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Switzerland, Thailand, T®rkiye, United Kingdom, Utd.Arab.Emir., Uzbekistan, Vietnam.
Lote
Model UDI-DI: 782115 (01)00884838099043(21); 782101 (01)00884838098275(21); 782140 (01)00884838108646(21) Serial Numbers: 45601 61291 61312 61315 61328 61336 61337 61339 61348 61349 61351 61354 61355 61356 61357 61359 61360 61361 61362 61363 61364 61367 61371 61373 61374 61375 61376 61377 61378 61380 61383 61384 61394 61395 61397 61403 61404 61406 61407 61412 61415 61416 87645 87648 87649 87651 87654 87655
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Ingenia 1.5T -For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 1) 782115 2) 782101 3) 782140 fue retirado del mercado?

Sí. FDA publicó el retiro Z-3194-2024 el 2 de octubre de 2024. Peligro reportado: Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter