Foley Catheter with temperature sensor 400TM 8FR, Catalogue Number 102201100880TY Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.
Published September 26, 2018 · Recall initiated: June 21, 2018
Summary
FDA records medical device Class II recall Z-2763-2018, published September 26, 2018 and initiated June 21, 2018. Recalling firm: Degania Silicone, Ltd.; Status: Terminated; Reported quantity: 4332; Lot: S17002877, S17002878, S17003909, S17004477, S17005607, S17006814, S17005115. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Degania Silicone, Ltd.
- Distributor
- Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
- Lot
- S17002877, S17002878, S17003909, S17004477, S17005607, S17006814, S17005115
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Israel
Related recalls
- FDACardinal Health Jackson-Pratt Channel Drain 19FR, 3/4 flutes, with trocar Cardinal Health: PN JP-2225 AnAugust 21, 2019
- FDAFoley Catheter with temperature sensor 400TM 16FR, Catalogue Number 102201101680TY Product Usage: This isSeptember 26, 2018
- FDATemperature Sensing 100% Silicone Foley Catheter, sterile, Catalogue Number 102201100880MD Product Usage: ThisSeptember 26, 2018
- FDATemperature sensor Foley catheter 10FR, Catalogue Number 102205101080AB Product Usage: This is sterile, singleSeptember 26, 2018
- FDATemperature Sensor Catheter 16FR, Catalogue Number 102201101663BI Product Usage: This is sterile, single use,September 26, 2018
- FDASesame Noodle Salad Vegan, packaged in 8 oz. or 16 oz. plastic container. UPC of 8 oz. container: 0 211630September 26, 2018
Questions about this recall
Was Foley Catheter with temperature sensor 400TM 8FR, Catalogue Number 102201100880TY Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period. recalled?
Yes. FDA published recall Z-2763-2018 on September 26, 2018. Reported hazard: The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
What should I do if I have this product?
Voluntary: Firm initiated · Letter