Monarch Platform (Endoscopy), REF: MON-000005-01, MON-000005-01R. Monarch Platform (Bronchoscopy), REF: MON-000006, MON-000006-RFB.
Publicado el 16 de octubre de 2024 · Inicio del retiro: 2 de agosto de 2024
Resumen
FDA registra el retiro Z-0051-2025 de dispositivos médicos de Clase II, publicado el 16 de octubre de 2024 e iniciado el 2 de agosto de 2024. Empresa responsable: AURIS HEALTH INC; Estado: En curso; Cantidad reportada: 176; Lote: Cart Power Cord 20' 14AWG BLK, PN 304-003613-00. Tower Power Cord 20' 14AWG BLK, PN 304-003607-00. REF/UDI-DI/Serial Numbers: MON-000005-01/+B634MON000005010/110024, 13, 110014, 110016, 110040, 110038, 110027, 110035, 110022, 110034, 1806P2, 120170, 110045, 110039, 110049, 110047, 110052, 110030, 110032, 110021, 110043, 110050, 110029, 110033, 110046, 110048, 110018, 110015, 110051, 1806P1, 110012, 14, 110041, 110036; MON-000005-01R/+B634MON00000501R0/110020, 110017, 110031, 110028, 120042, 110013. MON-000006/+B634MON0000060/120088, 120057, 120046, 120058, 120121, 120152, 120015, 120160, 120068, 120167, 120025, 120030, 120044, 120093, 120040, 120063, 120092, 120173, 120036, 120064, 120122, 120056, 120128, 120017, 120089, 120075, 120031, 120169, 120113, 120168, 120055, 120148, 120014, 120060, 120124, 120164, 120123, 120007, 120006, 120004, 120120, 120126, 120082, 120047, 120051, 120020, 120103, 120153, 120108, 120037, 120074, 120005, 120130, 120095, 120096, 120009, 120111, 120022, 120054, 120163, 120027, 120090, 120141, 120072, 120155, 120101, 120143, 120125, 120052, 120100, 120050, 120098, 120045, 120131, 120013, 120105, 120165, 120136, 120029, 120041, 120139, 120137, 120053, 120028, 120034, 120085, 120019, 120049, 120209, 120144, 120161, 120001, 120145, 120127, 120011, 120070, 120076, 120134, 120117, 120077, 120149, 120102, 120002, 120010, 120003, 120023, 120146, 120043, 120071, 120107, 120065, 120116, 120150, 120066, 120166, 120033, 120079, 120008, 120132, 120086, 120032, 120073, 120016, 120080, 120078, 120012, 120084, 120083, 120038, 120067, 120021; MON-000006-RFB/+B634MON0000060/120081, 120026, 120091, 120106, 120104. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Endoscopy/bronchoscopy/urology devices may have a strain relief defect with some cart and tower power cords that may cause an electrical short, which may lead to system shutdown, which can lead to prolonged procedure or case conversion, and if the user (e.g., operating room staff) touches exposed wires or conductive fluids in contact with the exposed wires, there is a chance of electrical shock.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- AURIS HEALTH INC
- Distribuidor
- Worldwide - US Nationwide distribution including in the states of MN, FL, NC, AK, NJ, NY, VA, WI, AZ, MS, LA, MI, MA, MT, WV, GA, DE, TX, CA, OH, IN, SC, PA, CT, ND, MO, DC, IL, NE, ID, WA, TN, OR, AL, SD, CO, KY, NM, VT, MD and the countries of Canada.
- Lote
- Cart Power Cord 20' 14AWG BLK, PN 304-003613-00. Tower Power Cord 20' 14AWG BLK, PN 304-003607-00. REF/UDI-DI/Serial Numbers: MON-000005-01/+B634MON000005010/110024, 13, 110014, 110016, 110040, 110038, 110027, 110035, 110022, 110034, 1806P2, 120170, 110045, 110039, 110049, 110047, 110052, 110030, 110032, 110021, 110043, 110050, 110029, 110033, 110046, 110048, 110018, 110015, 110051, 1806P1, 110012, 14, 110041, 110036; MON-000005-01R/+B634MON00000501R0/110020, 110017, 110031, 110028, 120042, 110013. MON-000006/+B634MON0000060/120088, 120057, 120046, 120058, 120121, 120152, 120015, 120160, 120068, 120167, 120025, 120030, 120044, 120093, 120040, 120063, 120092, 120173, 120036, 120064, 120122, 120056, 120128, 120017, 120089, 120075, 120031, 120169, 120113, 120168, 120055, 120148, 120014, 120060, 120124, 120164, 120123, 120007, 120006, 120004, 120120, 120126, 120082, 120047, 120051, 120020, 120103, 120153, 120108, 120037, 120074, 120005, 120130, 120095, 120096, 120009, 120111, 120022, 120054, 120163, 120027, 120090, 120141, 120072, 120155, 120101, 120143, 120125, 120052, 120100, 120050, 120098, 120045, 120131, 120013, 120105, 120165, 120136, 120029, 120041, 120139, 120137, 120053, 120028, 120034, 120085, 120019, 120049, 120209, 120144, 120161, 120001, 120145, 120127, 120011, 120070, 120076, 120134, 120117, 120077, 120149, 120102, 120002, 120010, 120003, 120023, 120146, 120043, 120071, 120107, 120065, 120116, 120150, 120066, 120166, 120033, 120079, 120008, 120132, 120086, 120032, 120073, 120016, 120080, 120078, 120012, 120084, 120083, 120038, 120067, 120021; MON-000006-RFB/+B634MON0000060/120081, 120026, 120091, 120106, 120104.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAMonarch Platform (Urology), REF: MON-000007, MON-000007-RFB16 de octubre de 2024
- FDAStealthStation S8 App versions 1.3.0 and 1.3.2, (Model # 9735762) The StealthStation System, with16 de octubre de 2024
- FDACheck-Flo Performer Introducer Set, Reference Part Number RCF-5.0-35-45-J, Order Number G27242. Performer16 de octubre de 2024
- FDAMedline Sterile Laparoscopy Pack, DYNJ68187A (REF), containing the component ConMed/Unimax Detachable Endo16 de octubre de 2024
- FDAMAKO Integrated Cutting System (MICS) HANDPIECE, Catalog Number 20906316 de octubre de 2024
- FDAMonarch Platform REF MON-000006 100-240V- 50/60Hz 1440w13 de diciembre de 2023
Preguntas sobre este retiro
¿Monarch Platform (Endoscopy), REF: MON-000005-01, MON-000005-01R. Monarch Platform (Bronchoscopy), REF: MON-000006, MON-000006-RFB. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0051-2025 el 16 de octubre de 2024. Peligro reportado: Endoscopy/bronchoscopy/urology devices may have a strain relief defect with some cart and tower power cords that may cause an electrical short, which may lead to system shutdown, which can lead to prolonged procedure or case conversion, and if the user (e.g., operating room staff) touches exposed wires or conductive fluids in contact with the exposed wires, there is a chance of electrical shock.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter