Hemorrhoidal Circular Stapler and Accessories 33mm Product Usage: The PROXIMATE HCS Hemorrhoidal Circular Staplers, (PPH01 and PPH03), and accessories have application for general surgical treatment of anorectal wall defects by means of transanal stapling and resection of mucosal and musculo mucosal tissue. The TRANSTARTM CIRCULAR STAPLER, (STR10), and accessories have application for general surgical treatment of anorectal wall defects by means of transanal stapling and resection of mucosal and musculomucosal tissue.
Publicado el 24 de octubre de 2012 · Inicio del retiro: 3 de agosto de 2012
Resumen
FDA registra el retiro Z-0057-2013 de dispositivos médicos de Clase I, publicado el 24 de octubre de 2012 e iniciado el 3 de agosto de 2012. Empresa responsable: Ethicon Endo-Surgery Inc; Estado: Finalizado; Cantidad reportada: 157,881 devices; Lote: PPH01: 68 lots; 19,030 pieces; Expiration date April 2016 -June 2017; PPH03: 494 lots; 144,963 pieces; Expiration date March 2016 - June 2017; STR10: 3 lots; 432 pieces; Expiration date Sept 2016 - Oct 2016. La Clase I indica una probabilidad razonable de consecuencias graves para la salud o muerte.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Ethicon Endo-Surgery initiated this voluntary global recall for specific production lots of PROXIMATE® PPH Hemorrhoidal Circular Stapler and Accessories 33mm (PPH01 and PPH03) and Transtar® Circular Stapler Procedure Set (STR10) due to difficulty firing the device which may result in incomplete firing stroke that may result in an incomplete staple formation.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Ethicon Endo-Surgery Inc
- Distribuidor
- Worldwide Distribution - US (nationwide) in the states of: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the countries of: ARGENTINA, AUSTRALIA, BAHRAIN, BELGIUM, BRAZIL, CHILE, CHINA, COLOMBIA, COSTA RICA, CYPRUS, DOMINICAN REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, GREECE, GUATEMALA, HONG KONG, INDIA, INDONESIA, ISRAEL, ITALY, JAPAN, JORDAN, REPUBLIC OF KOREA, KUWAIT, LEBANON, MALAYSIA, MEXICO, NEPAL, NICARAGUA, PAKISTAN, PANAMA, PARAGUAY, PERU, PHILIPPINES, PUERTO RICO, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIRATES, URUGUAY, VENEZUELA, VIETNAM, and YEMEN. There was one Canadian consignee and 43 U.S. Government consignees.
- Lote
- PPH01: 68 lots; 19,030 pieces; Expiration date April 2016 -June 2017; PPH03: 494 lots; 144,963 pieces; Expiration date March 2016 - June 2017; STR10: 3 lots; 432 pieces; Expiration date Sept 2016 - Oct 2016
- Clasificación de riesgo
- Class I
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAECHELON LINEAR Cutters Reload 80 mm Blue, intended for transection, resection, and/or creation of anastomoses,22 de octubre de 2025
- FDAENDOPATH ECHELON 45mm RELOADS WITH GRIPPING SURFACE TECHNOLOGY (GST), ECHELON Endopath Stapler, Endoscopic7 de febrero de 2024
- FDALigamax 5 mm Endoscopic Multiple Clip Applier18 de diciembre de 2019
- FDAECHELON FLEX Powered Plus ENDOPATH 60mm Stapler - 280mm shaft, Product Code PCEE60A6 de noviembre de 2019
- FDAAutoMate 1250 and AutoMate 2550, Part Numbers ODL25125 and ODL25255. The AutoMate 1200/2500 System Series is a24 de octubre de 2012
- FDABelmont Rapid Infuser, a Fluid Management System, Model FMS2000.24 de octubre de 2012
Preguntas sobre este retiro
¿Hemorrhoidal Circular Stapler and Accessories 33mm Product Usage: The PROXIMATE HCS Hemorrhoidal Circular Staplers, (PPH01 and PPH03), and accessories have application for general surgical treatment of anorectal wall defects by means of transanal stapling and resection of mucosal and musculo mucosal tissue. The TRANSTARTM CIRCULAR STAPLER, (STR10), and accessories have application for general surgical treatment of anorectal wall defects by means of transanal stapling and resection of mucosal and musculomucosal tissue. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0057-2013 el 24 de octubre de 2012. Peligro reportado: Ethicon Endo-Surgery initiated this voluntary global recall for specific production lots of PROXIMATE® PPH Hemorrhoidal Circular Stapler and Accessories 33mm (PPH01 and PPH03) and Transtar® Circular Stapler Procedure Set (STR10) due to difficulty firing the device which may result in incomplete firing stroke that may result in an incomplete staple formation.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter