Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0057-2013

Hemorrhoidal Circular Stapler and Accessories 33mm Product Usage: The PROXIMATE HCS Hemorrhoidal Circular Staplers, (PPH01 and PPH03), and accessories have application for general surgical treatment of anorectal wall defects by means of transanal stapling and resection of mucosal and musculo mucosal tissue. The TRANSTARTM CIRCULAR STAPLER, (STR10), and accessories have application for general surgical treatment of anorectal wall defects by means of transanal stapling and resection of mucosal and musculomucosal tissue.

Publicado el 24 de octubre de 2012 · Inicio del retiro: 3 de agosto de 2012

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Ethicon Endo-Surgery initiated this voluntary global recall for specific production lots of PROXIMATE¿ PPH Hemorrhoidal Circular Stapler and Accessories 33mm (PPH01 and PPH03) and Transtar¿ Circular Stapler Procedure Set (STR10) due to difficulty firing the device which may result in incomplete firing stroke that may result in an incomplete staple formation.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Ethicon Endo-Surgery Inc
Distribuidor
Worldwide Distribution - US (nationwide) in the states of: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the countries of: ARGENTINA, AUSTRALIA, BAHRAIN, BELGIUM, BRAZIL, CHILE, CHINA, COLOMBIA, COSTA RICA, CYPRUS, DOMINICAN REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, GREECE, GUATEMALA, HONG KONG, INDIA, INDONESIA, ISRAEL, ITALY, JAPAN, JORDAN, REPUBLIC OF KOREA, KUWAIT, LEBANON, MALAYSIA, MEXICO, NEPAL, NICARAGUA, PAKISTAN, PANAMA, PARAGUAY, PERU, PHILIPPINES, PUERTO RICO, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIRATES, URUGUAY, VENEZUELA, VIETNAM, and YEMEN. There was one Canadian consignee and 43 U.S. Government consignees.
Lote
PPH01: 68 lots; 19,030 pieces; Expiration date April 2016 -June 2017; PPH03: 494 lots; 144,963 pieces; Expiration date March 2016 - June 2017; STR10: 3 lots; 432 pieces; Expiration date Sept 2016 - Oct 2016
Clasificación de riesgo
Class I
Área (FDA)
device
País
United States

¿Vendes productos como este?