Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0117-2013

HiRes 90K devices with Helix Electrodes Model CI 1400-02H The HiResolution Bionic Ear System is a cochlear implant system designed to provide useful hearing to individuals with severe to profound hearing loss. It consists of internal and external components. The internal components include a stimulator (HiRes90K ICS) and electrode array (HiFocus 1j or Helix). The HiRes90K ICS is implanted surgically under the skin behind the ear. The electrode array is inserted into the cochlea. The external components include a sound processor (body-worn or ear-level), a headpiece, and a cable. The system converts sound into electrical energy that activates the auditory nerve. The auditory nerve then sends information to the brain, where it is interpreted as sound. The Helix consists of a fantail, electrode lead, and Helix electrode array. The electrode, comprised of platinum-iridium wires, is housed in a silicone carrier and extends from the titanium case. The Helix electrode array is 24.5 mm in length and is designed to be inserted approximately 18-21 mm into a normal patent cochlea. It consists of 16 planar contacts arranged along the medial surface of the electrode array for stimulation of discrete segments of the cochlea. The Helix electrode insertion tool is designed to assist the surgeon with the insertion of the Helix electrode. Only the Helix insertion should be used for insertion of the Helix electrode; the use of other instruments may result in damage to the electrode. The insertion tool is a single-use instrument.

Publicado el 31 de octubre de 2012 · Inicio del retiro: 14 de septiembre de 2007

Resumen

FDA registra el retiro Z-0117-2013 de dispositivos médicos de Clase II, publicado el 31 de octubre de 2012 e iniciado el 14 de septiembre de 2007. Empresa responsable: Advanced Bionics Corporation; Estado: Finalizado; Cantidad reportada: 17 units; Lote: Serial Numbers: 351323, 251153, 350989, 351226, 350196, 351136, 350140, 350201, 351261, 351264, 351266, 351269, 351412, 351415, 351429, 350157, 351303. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Advanced Bionics is recalling the the HiRes 90K cochlear implant with HiFocus Helix Electrode because of loose-fitting insertion tool assemblies.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Advanced Bionics Corporation
Distribuidor
Worldwide distribution: USA (nationwide) including state: IL and country of: Europe.
Lote
Serial Numbers: 351323, 251153, 350989, 351226, 350196, 351136, 350140, 350201, 351261, 351264, 351266, 351269, 351412, 351415, 351429, 350157, 351303.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿HiRes 90K devices with Helix Electrodes Model CI 1400-02H The HiResolution Bionic Ear System is a cochlear implant system designed to provide useful hearing to individuals with severe to profound hearing loss. It consists of internal and external components. The internal components include a stimulator (HiRes90K ICS) and electrode array (HiFocus 1j or Helix). The HiRes90K ICS is implanted surgically under the skin behind the ear. The electrode array is inserted into the cochlea. The external components include a sound processor (body-worn or ear-level), a headpiece, and a cable. The system converts sound into electrical energy that activates the auditory nerve. The auditory nerve then sends information to the brain, where it is interpreted as sound. The Helix consists of a fantail, electrode lead, and Helix electrode array. The electrode, comprised of platinum-iridium wires, is housed in a silicone carrier and extends from the titanium case. The Helix electrode array is 24.5 mm in length and is designed to be inserted approximately 18-21 mm into a normal patent cochlea. It consists of 16 planar contacts arranged along the medial surface of the electrode array for stimulation of discrete segments of the cochlea. The Helix electrode insertion tool is designed to assist the surgeon with the insertion of the Helix electrode. Only the Helix insertion should be used for insertion of the Helix electrode; the use of other instruments may result in damage to the electrode. The insertion tool is a single-use instrument. fue retirado del mercado?

Sí. FDA publicó el retiro Z-0117-2013 el 31 de octubre de 2012. Peligro reportado: Advanced Bionics is recalling the the HiRes 90K cochlear implant with HiFocus Helix Electrode because of loose-fitting insertion tool assemblies.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter