FDAFDA · Alimentos y medicamentos·N.º oficial Z-0211-2016
RotaWire Elite and wireClip Torquer, Guidewire and Guidewire Manipulation Device, Floppy. REF H802223301 (5-pack outer package), REF H802223300 (single unit). The Rotablator rotational atherectomy system consists of a diamond tipped burr capable of high speed rotation designed for plaque modification and debulking in vascular interventions. This system is designed to travel over a specialized 0.009 guidewire to guide the path of burr travel to deliver therapy where intended. The guidewires are intended for use with the Rotablator Rotational Atherectomy System.
Publicado el 18 de noviembre de 2015 · Inicio del retiro: 9 de octubre de 2015
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Boston Scientific is recalling its recently released RotaWire Elite Guidewire and wireClip Torquer Guidewire Manipulation device that is used in conjunction with the Rotablator Rotational Atherectomy System because of complaints of wire fracture.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Boston Scientific Corporation
- Distribuidor
- Worldwide Distribution: US Distribution in the states of AZ, CA, DE, FL, GA, IL, LA, MD, MI, MN, NE, NY, PA, SC, TX, WA, and WI, and the countries of Belgium, Germany, Great Britain, Italy, Japan, and South Africa.
- Lote
- Lots: 18201519, 18212726, 18256610, 18261218, 18271412, 18282032, 18362094, 18378161, 18379210, 18384443
- Clasificación de riesgo
- Class I
- Área (FDA)
- device
- País
- United States