A1CNow+ Multi-Test HbA1c System consists of (1) a semi-disposable plastic-encased device (the monitor), (2) a plastic cartridge enclosing dry reagent strips, and (3) a sample dilution kit for: collecting the blood sample, mixing the sample with the required pre-treatment solution, and delivering the sample to the cartridge.

Publicado el 30 de octubre de 2024 · Inicio del retiro: 18 de septiembre de 2024

Estado reportado al momento de publicación: Completed — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The reason for this recall is PTS Diagnostics, Inc. received complaints of non-matching cartridge and analyzer codes within the product test system box. Use of cartridges and analyzers with non-matching codes may result in inaccurate test results as components within each lot number are meant to be used together, and not mixed between lots. Therefore, the test results may possibly lead to Hemoglobin A1C positive or negative bias results. While the product is for monitoring only, over recovery with a higher than accurate value can be an indication for increased therapy; such as physical activity, weight reduction, meal management, pharmaceutical therapy.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Polymer Technology Systems, Inc. dba PTS Diagnostics, Inc.
Distribuidor
Worldwide distribution - US Nationwide and the countries of Japan, South Africa.
Lote
UDI-DI: 00381933021011 Lot Numbers: 9331803, 2323019
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States