Hoffman LRF Telescopic Strut Short Stryker Stryker Trauma AG CH-2545 Seizach Distributed in the USA by Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430. The use of external devices provides the surgeon with a means of bone fixation for the management of fracture and reconstructive surgery. These devices are intended only to assist healing and are not intended to replace normal bone structures. External fixation devices are intended to hold fractured bone surfaces in apposition to facilitate normal healing. While proven successful, the devices are manufactured from metal, plastic, and composite materials. No such fracture fixation device, subject to fatigue, can be expected to withstand activity levels in the same way as would a normal health bone. The external fixation system, therefore, will not be as strong, reliable or durable as a normal human bone.
Publicado el 14 de noviembre de 2012 · Inicio del retiro: 3 de julio de 2012
Resumen
FDA registra el retiro Z-0266-2013 de dispositivos médicos de Clase II, publicado el 14 de noviembre de 2012 e iniciado el 3 de julio de 2012. Empresa responsable: Stryker Howmedica Osteonics Corp.; Estado: Finalizado; Cantidad reportada: 165 units; Lote: K113327 Catalog No: Telescopic Strut Extra Short Hoffman LRF Length: 4933-0-100 :100-125mm (Black) 4933-0-120 Telescopic Strut Short Hoffman LRF Length: 119-161 mm (Yellow) 4933-0-140 Telescopic Strut Medium Hoffman LRF Length 138-201 mm (Blue) 4933-0-180 Telescopic Strut Long Hoffman LRF Length: 177-277 mm (Red); Lot Code: 4933-0-100 Telescopic Strut Extra Short Hoffman LRF Length 100-125mm (Black) X28937; Z04197; Z09163; Z09597; Z12135; 4933-0-120 Telescopic Strut Short Hoffman LRF Length: 119-161 mm (Yellow) X28961; Z04199; Z09161; Z09769; Z12012; 4933-0-140 Telescopic Strut Medium Hoffman LRF Length 138-201 mm (Blue) X34352; Z04200; Z05068; Z09011; Z09598; Z10872; 4933-0-180 Telescopic Strut Long Hoffman LRF Length: 177-277 mm (Red); X28963 ;Z04429; Z09471; Z11686. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Three complaints have been filed where the Hoffman LRF Telescopic Struts have broken during load bearing application by patients having a body weight greater than 250 LBS.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Stryker Howmedica Osteonics Corp.
- Distribuidor
- Nationwide Distribution.
- Lote
- K113327 Catalog No: Telescopic Strut Extra Short Hoffman LRF Length: 4933-0-100 :100-125mm (Black) 4933-0-120 Telescopic Strut Short Hoffman LRF Length: 119-161 mm (Yellow) 4933-0-140 Telescopic Strut Medium Hoffman LRF Length 138-201 mm (Blue) 4933-0-180 Telescopic Strut Long Hoffman LRF Length: 177-277 mm (Red); Lot Code: 4933-0-100 Telescopic Strut Extra Short Hoffman LRF Length 100-125mm (Black) X28937; Z04197; Z09163; Z09597; Z12135; 4933-0-120 Telescopic Strut Short Hoffman LRF Length: 119-161 mm (Yellow) X28961; Z04199; Z09161; Z09769; Z12012; 4933-0-140 Telescopic Strut Medium Hoffman LRF Length 138-201 mm (Blue) X34352; Z04200; Z05068; Z09011; Z09598; Z10872; 4933-0-180 Telescopic Strut Long Hoffman LRF Length: 177-277 mm (Red); X28963 ;Z04429; Z09471; Z11686
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAHoffmann LRF Safety Clip (Bone Transport Strut), 14 x 12 x 5 mm, REF 4933-0-401, NDC31 de mayo de 2017
- FDASupracondylar Nail T2 SCN 09x300 mm T2 Supracondylar Nail System is intended to provide strong and stable21 de septiembre de 2016
- FDAStryker Orthopaedics Specialty Triathlon Tibial Alignment Handle with Secondary Lock, Non-Sterile, Used to17 de agosto de 2016
- FDAStryker, T2 Guidewire, Ball-Tipped, 3 x 1000mm Product Usage: Intended for the temporary stabilization of bone2 de marzo de 2016
- FDA0.4/0.7JG326D-265AX/AT Product Usage: The device is a mobile X-ray system in order to take general radiography14 de noviembre de 2012
- FDAQuadra S Offset Broach Handle The offset broach handle is intended to attach to the broaches of different14 de noviembre de 2012
Preguntas sobre este retiro
¿Hoffman LRF Telescopic Strut Short Stryker Stryker Trauma AG CH-2545 Seizach Distributed in the USA by Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430. The use of external devices provides the surgeon with a means of bone fixation for the management of fracture and reconstructive surgery. These devices are intended only to assist healing and are not intended to replace normal bone structures. External fixation devices are intended to hold fractured bone surfaces in apposition to facilitate normal healing. While proven successful, the devices are manufactured from metal, plastic, and composite materials. No such fracture fixation device, subject to fatigue, can be expected to withstand activity levels in the same way as would a normal health bone. The external fixation system, therefore, will not be as strong, reliable or durable as a normal human bone. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0266-2013 el 14 de noviembre de 2012. Peligro reportado: Three complaints have been filed where the Hoffman LRF Telescopic Struts have broken during load bearing application by patients having a body weight greater than 250 LBS.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter