Spectra System/Legacy Guided Surgery Handle Kit (Part Number: G-HK) Spectra System/Legacy Guided Surgery Handle Kit (Part Number: G-HK) allows clinicians to drill through a surgical guide to create an osteotomy for implant placement with the precision and accuracy as designed with treatment planning software. The G-HK consists of 4 guide handles with inserts attached on each end. Each insert has an opening that matches the diameter of our surgical drills. The insert portion of the guide handle fits intimately into the guide sleeve that is embedded into a surgical guide.
Publicado el 27 de diciembre de 2017 · Inicio del retiro: 15 de noviembre de 2017
Resumen
FDA registra el retiro Z-0283-2018 de dispositivos médicos de Clase II, publicado el 27 de diciembre de 2017 e iniciado el 15 de noviembre de 2017. Empresa responsable: Implant Direct Sybron Manufacturing, LLC; Estado: Finalizado; Cantidad reportada: 560 units (worldwide); Lote: Lot # 38665, 38825, 42562, 42649, 43810, 45014, 46950, 50165, 45018, 50165, 52360, 52987, 60376, 62435, 64245, 67044, 73431, 73729, 74772, 70293. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The Spectra System/Legacy Guided Surgery Handle Kit had guided handle inserts that were out of Implant Direct specifications, from the results of complaints submitted by clinicians regarding the 3.4mm guide insert that will not allow the 3.4 drill to pass through, since the diameter is too small. In addition, the 2.3mm guide was too large.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Implant Direct Sybron Manufacturing, LLC
- Distribuidor
- Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Europe, Canada, Australia, Egypt, Jordan, and Lebanon.
- Lote
- Lot # 38665, 38825, 42562, 42649, 43810, 45014, 46950, 50165, 45018, 50165, 52360, 52987, 60376, 62435, 64245, 67044, 73431, 73729, 74772, 70293
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAImplantDirect InterActive Precision IO Scan Adapter, Part #6530-09PT, Platform: 3.0mm, Non-Sterile, Rx. The15 de enero de 2020
- FDAImplantDirect InterActive Precision IO Scan Adapter, Part #6534-09PT, Platform: 3.4mm, Non-Sterile, Rx. The15 de enero de 2020
- FDAImplant Direct, REF654311U, simply InterActive Implant, Surface: SBM, Qty 1, 4.3mmD x 11.5mmL, 3.4mmD20 de noviembre de 2019
- FDAImplantDirect, Dental implants intended for use in partially and fully edentulous upper and lower jaws in1 de agosto de 2018
- FDAZimmer Dental Tapered SwissPlus & SwissPlus Implant Systems surgical kit, Ref OPCST. The label shows the27 de diciembre de 2017
- FDATapered Screw-Vent MTX Model TSVTB1127 de diciembre de 2017
Preguntas sobre este retiro
¿Spectra System/Legacy Guided Surgery Handle Kit (Part Number: G-HK) Spectra System/Legacy Guided Surgery Handle Kit (Part Number: G-HK) allows clinicians to drill through a surgical guide to create an osteotomy for implant placement with the precision and accuracy as designed with treatment planning software. The G-HK consists of 4 guide handles with inserts attached on each end. Each insert has an opening that matches the diameter of our surgical drills. The insert portion of the guide handle fits intimately into the guide sleeve that is embedded into a surgical guide. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0283-2018 el 27 de diciembre de 2017. Peligro reportado: The Spectra System/Legacy Guided Surgery Handle Kit had guided handle inserts that were out of Implant Direct specifications, from the results of complaints submitted by clinicians regarding the 3.4mm guide insert that will not allow the 3.4 drill to pass through, since the diameter is too small. In addition, the 2.3mm guide was too large.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter