LANX Timberline MPF Lateral Modular Plate Fixation System Straight Split Tip Driver - Short ("Driver"). Model Number 8630-0107. The Timberline MPF device is an intervertebral body fusion device consisting of a PEEK intervertebral spacer, titanium plate, and screws. The PEEK spacer has a generally rounded shape with various heights and footprints and has a hollowed out central area to accommodate autogenous bone graft. The upper and lower surfaces have a series of transverse grooves formed to improve stability and fixation once the device is inserted. The titanium plate has holes for receiving bone screws and a central hole for receiving a cover plate to prevent screw back-out. The Timberline MPF system is available in a variety of sizes and configurations to approximate anatomical variation in different vertebral levels and/or patient anatomy. The Timberline MPF system is provided nonsterile. The Driver is used with the Timberline MPF system to drive titanium screws through the plate and into the vertebrae thus locking plate to the vertebrae.
Publicado el 4 de diciembre de 2013 · Inicio del retiro: 4 de noviembre de 2013
Resumen
FDA registra el retiro Z-0400-2014 de dispositivos médicos de Clase II, publicado el 4 de diciembre de 2013 e iniciado el 4 de noviembre de 2013. Empresa responsable: Lanx, Inc.; Estado: Finalizado; Cantidad reportada: 25 units; Lote: Lot No: TU00090, TU00157 and TU00242. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Lanx, Inc. is recalling various lots of Timberline MPF Straight Split Tip Driver - Short ("Driver") because the width of the most distal end from the Driver handle (the tip of the hex portion of the Driver) is oversized.
Qué hacer
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Lanx, Inc.
- Distribuidor
- Nationwide Distribution including AZ, PA, WI, NY, TX, PR, and TN.
- Lote
- Lot No: TU00090, TU00157 and TU00242
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAMedtronic Strata II Shunt Assembly Kit, Small, Catalog No. 27848. Strata II Shunt Assemblies are designed to4 de diciembre de 2013
- FDASynthes 4.5MM TI VA-LCP CRVD Condylar Plate/6H/159MM/RT-STER Synthes 4.5MM TI VA-LCP CRVD Condylar4 de diciembre de 2013
- FDAMedtronic Delta Shunt Assembly Kit with Bioglide Small, Performance Level 1.0, Catalog No. 276201. Delta Shunt4 de diciembre de 2013
- FDASTERRAD 100NX, Product Code: 10104 The STERRAD Steri lization System is a low temperature general purpose4 de diciembre de 2013
- FDACSF-Unitized Shunt, Contoured Regular, Low Pressure, Catalog No. 46022; CSF-Unitized Shunt, Contoured Regular,4 de diciembre de 2013
- FDASTERRAD NX, Product Code: 10033 The STERRAD Steri lization System is a low temperature general purpose4 de diciembre de 2013
Preguntas sobre este retiro
¿LANX Timberline MPF Lateral Modular Plate Fixation System Straight Split Tip Driver - Short ("Driver"). Model Number 8630-0107. The Timberline MPF device is an intervertebral body fusion device consisting of a PEEK intervertebral spacer, titanium plate, and screws. The PEEK spacer has a generally rounded shape with various heights and footprints and has a hollowed out central area to accommodate autogenous bone graft. The upper and lower surfaces have a series of transverse grooves formed to improve stability and fixation once the device is inserted. The titanium plate has holes for receiving bone screws and a central hole for receiving a cover plate to prevent screw back-out. The Timberline MPF system is available in a variety of sizes and configurations to approximate anatomical variation in different vertebral levels and/or patient anatomy. The Timberline MPF system is provided nonsterile. The Driver is used with the Timberline MPF system to drive titanium screws through the plate and into the vertebrae thus locking plate to the vertebrae. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0400-2014 el 4 de diciembre de 2013. Peligro reportado: Lanx, Inc. is recalling various lots of Timberline MPF Straight Split Tip Driver - Short ("Driver") because the width of the most distal end from the Driver handle (the tip of the hex portion of the Driver) is oversized.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit