Cutera brand truSculpt infrared System (optional), Massager, Vacuum, radio frequency induced heat, Product is manufactured and distributed by Cutera, Inc., Brisbane, CA The truSculpt massage device is intended to provide a temporary reduction in the appearance of cellulite. The truSculpt infrared (optional) and RF energies are intended to provide topical heating for the purpose of elevating tissue temperature for the treatment of selected medical conditions such as relief of pain, muscle spasms, and increase in local circulation.
Publicado el 18 de diciembre de 2013 · Inicio del retiro: 16 de octubre de 2013
Resumen
FDA registra el retiro Z-0427-2014 de dispositivos médicos de Clase II, publicado el 18 de diciembre de 2013 e iniciado el 16 de octubre de 2013. Empresa responsable: Cutera, Inc.; Estado: Finalizado; Cantidad reportada: 451; Lote: Serial Numbers: TS10140 TS10052 TS10107 TS10161 TS10086 TS10136 TS10083 TS10045 TS10032 TS10075 TS10100 TS10070 TS10129 TS10049 TS10040 TS10055 TS10160 TS10046 TS10117 TS10137 TS10141 TS10080 TS10097 TS10037 TS10131 TS10138 TS10119 TS10088 TS10074 TS10114 TS10145 TS10066 TS10106 TS10155 TS10056 TS10094 TS10139 TS10022 TS10128 TS10112 TS10068 TS10124 TS10078 TS10134 TS10079 TS10002 TS10109 TS10156 TS10135 TS10110 TS10008 TS10115 TS10093 TS10146 TS10050 TS10017 TS10054 TS10029 TS10104 TS10048 TS10090 TS10062 TS10024 TS10158 TS10071 TS10035 TS10059 TS10153 TS10025 TS10087 TS10130 TS10133 TS10101 TS10103 TS10125 TS10018 TS10089 TS10023 TS10147 TS10033 TS10015 TS10014 TS10039 TS10085. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Cutera received two adverse events associated with the edge of the electrode, when it was pressed against neck skin over the thyroid cartilage and due to the absence of adequate underlying conductive tissue, increased skin heating occurred. Both patients developed linear welts and blisters consistent with partial thickness thermal burns to the skin.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Cutera, Inc.
- Distribuidor
- Worldwide Distribution - USA (nationwide) and Internationally to Canada
- Lote
- Serial Numbers: TS10140 TS10052 TS10107 TS10161 TS10086 TS10136 TS10083 TS10045 TS10032 TS10075 TS10100 TS10070 TS10129 TS10049 TS10040 TS10055 TS10160 TS10046 TS10117 TS10137 TS10141 TS10080 TS10097 TS10037 TS10131 TS10138 TS10119 TS10088 TS10074 TS10114 TS10145 TS10066 TS10106 TS10155 TS10056 TS10094 TS10139 TS10022 TS10128 TS10112 TS10068 TS10124 TS10078 TS10134 TS10079 TS10002 TS10109 TS10156 TS10135 TS10110 TS10008 TS10115 TS10093 TS10146 TS10050 TS10017 TS10054 TS10029 TS10104 TS10048 TS10090 TS10062 TS10024 TS10158 TS10071 TS10035 TS10059 TS10153 TS10025 TS10087 TS10130 TS10133 TS10101 TS10103 TS10125 TS10018 TS10089 TS10023 TS10147 TS10033 TS10015 TS10014 TS10039 TS10085
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAVitacost Super Daily Enzymes 500 mg caps, product is packaged in a plastic bottle with 240 capsules per18 de diciembre de 2013
- FDAiba COMPASS SW Version 3.0, Catalog Number CS10-100 (medical linear accelerator) radiological18 de diciembre de 2013
- FDADigestive Enzyme Complex/Enzyme Formula Caps, Quantity:13M, Product #: IB286, product is packaged in bulk18 de diciembre de 2013
- FDAProduct is labeled in part: LIST NO. 43449; RECEPTAL CANISTER; DO NOT FILL 900 ABOVE THIS LINE; Product of18 de diciembre de 2013
- FDAVitacost Mega Spectrum Enzyme, packaged with 180 Vegetarian Capsules, UPC Code: 84419701519. Manufactured for18 de diciembre de 2013
- FDAEthicon Stratafix Symmetric PDS Plus Knotless Tissue Control Device Ethicon LLC Guaynabo, Puerto Rico 00969.18 de diciembre de 2013
Preguntas sobre este retiro
¿Cutera brand truSculpt infrared System (optional), Massager, Vacuum, radio frequency induced heat, Product is manufactured and distributed by Cutera, Inc., Brisbane, CA The truSculpt massage device is intended to provide a temporary reduction in the appearance of cellulite. The truSculpt infrared (optional) and RF energies are intended to provide topical heating for the purpose of elevating tissue temperature for the treatment of selected medical conditions such as relief of pain, muscle spasms, and increase in local circulation. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0427-2014 el 18 de diciembre de 2013. Peligro reportado: Cutera received two adverse events associated with the edge of the electrode, when it was pressed against neck skin over the thyroid cartilage and due to the absence of adequate underlying conductive tissue, increased skin heating occurred. Both patients developed linear welts and blisters consistent with partial thickness thermal burns to the skin.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter