Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0483-2014

Stryker Orthopaedics Cancellous Bone Screw Howmedica Osteonics Corp. Stryker France Both the 6.4 mm diameter, 25mm length, Cancellous Bone Screws (Cat no 2030-6525-1) and the 5.5mm diameter, 40 mm length Cancellous Bone Screws (Cat no 2029-5540-1) are cleared for use in Total Hip Arthroplasty (THA), when implanting Stryker's Hemispherical Cluster Hole Acetabular Shell Systems and Total Knee Arthroplasty (TKA) when implanting Stryker's Series 7000 Standard Tibial Tray systems. The use of bone screws, during both THA and TKA are intended as supplemental fixation.

Publicado el 18 de diciembre de 2013 · Inicio del retiro: 1 de octubre de 2013

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Stryker received a report from the field indicating that a 5.5 mm Cancellous Bone Screw 40 mm was identified in packaging associated with a 6.5 mm Cancellous Bone Screw 25 mm.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Stryker Howmedica Osteonics Corp.
Distribuidor
Nationwide Distribution
Lote
1) 6.5 mm Cancellous Bone Screw 25mm Catalog No: 2030-6525-1 Lot code MMLNLA 2) 5.5 Cancellous Bone Screw 40 mm Catalog No: 2029-5540-1 Lot Code MMLMJ1
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States