Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0497-2025

Vue PACS Diagnostic Viewer with Vue PACS client 12.1.5 to 12.2.8.454 software Model Number: 1149418 Intended use is to provide completely scalable local and wide area PACS (Picture Archiving and Communication System) solutions for hospital and related institutions/sites.

Publicado el 27 de noviembre de 2024 · Inicio del retiro: 10 de octubre de 2024

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

A software issue affecting Philips Vue PACS may cause incorrect ischemic map and table value calculations when using the export function in the Perfusion Application for a Siemens CT perfusion study, resulting in misdiagnosis due to incorrect ischemic map and table values.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Philips Medical Systems Technologies Ltd.
Distribuidor
US Nationwide and Worldwide Distribution Foreign: Country Argentina Australia Austria Belgium Brazil Chile China Colombia Denmark Finland France Germany Gibraltar Greece Guernsey Holy See (Vatican City State) Hong Kong Hungary Indonesia Ireland Israel Italy Japan Jersey Jordan Lebanon Malaysia Mexico Netherlands New Zealand Norway Oman Panama Philippines Poland Portugal Romania Saudi Arabia Serbia Slovenia Spain Sweden Switzerland Taiwan Thailand United Kingdom
Lote
UDI-DI: (01)00884838100350(11)200221(10)12.2.5; (01)00884838100350(11)200812(10)12.2.6.0; (01)00884838100350(11)201207(10)12.2.6.100; (01)00884838100350(11)210303(10)12.2.6.200; (01)00884838100350(11)210518(10)12.2.6.201; (01)00884838100350(11)210729(10)12.2.6.300; (01)00884838100350(11)210502(10)12.2.8.0; (01)00884838100350(11)220112(10)12.2.8.100; (01)00884838100350(11)221213(10)12.2.8.200; (01)00884838100350(11)230301(10)12.2.8.210; (01)00884838100350(11)230320(10)12.2.8.300; (01)00884838100350(11)230608(10)12.2.8.310; (01)00884838100350(11)230718(10)12.2.8.400; (01)00884838100350(11)230810(10)12.2.8.401; (01)00884838100350(11)230824(10)12.2.8.317; (01)00884838100350(11)231019(10)12.2.8.412; (01)00884838100350(11)231108(10)12.2.8.403; (01)00884838100350(11)231129(10)12.2.8.413; (01)00884838100350(11)231214(10)12.2.8.414; (01)00884838100350(11)231204(10)12.2.8.420; (01)00884838100350(11)240101(10)12.2.8.415; (01)00884838100350(11)240116(10)12.2.8.421; (01)00884838100350(11)240204(10)12.2.8.422; (01)00884838100350(11)240226(10)12.2.8.424; (01)00884838100350(11)240311(10)12.2.8.426; (01)00884838100350(11)240407(10)12.2.8.440; (01)00884838100350(11)240425(10)12.2.8.450; (01)00884838100350(11)240613(10)12.2.8.451; (01)00884838100350(11)240715(10)12.2.8.452; (01)00884838100350(11)240715(10)12.2.8.454; (01)00884838100350(11)230926(10)12.2.8.410
Clasificación de riesgo
Class II
Área (FDA)
device
País
Israel

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta