FDAFDA · Alimentos y medicamentos·N.º oficial Z-0524-2015
Medtronic MiniMed Silhouette and/or MiniMed Sure-T Infusion Sets; Paradigm Silhouette infusion sets: MMT-368, MMT-368600, MMT-371, MMT-373, MMT-377, MMT-377600, MMT378, MMT-378600, MMT-38 1, MMT-381 600, MMT-382, MMT-382600, MMT-383, MMT-383600, MMT384, & MMT384600. Sure-T infusion sets MMT-862, MMT-863, MMT-864, MMT-865, MMT-866, MMT-873, MMT-874, MMTMMT-875, MMT-876, MMT-884, & MMT-886. MiniMed Duo sensor/infusion sets: MMT-5001A23, MMT-5001A43, MMT-5001B23, MMT-5001B43, MMT-5011A23, MMT-5011A43, MMT-5011B23, & MMT-5011B43. Silhouette and Sure-T disposable infusion sets are used with Medtronic insulin infusion pumps using Paradigm P-cap connectors. The infusion set connects the insulin reservoir in the pump to the patients body.
Publicado el 17 de diciembre de 2014 · Inicio del retiro: 4 de noviembre de 2014
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Tubing may detach at the connect/disconnect location on the MiniMed Silhouette and MiniMed Sure-T infusion sets. Detachment causes interrupted insulin delivery and the pump will not alarm to notify your patient. This can lead to hyperglycemia, and diabetic ketoacidosis (DKA).
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Medtronic MiniMed Inc.
- Distribuidor
- Worldwide distribution: US (Nationwide) and countries of: Algeria, Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Bermuda, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lativa, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malaysia, Mexico, Morocco, Namibia, Netherlands, New Zealand, Nicaragua, Norway, Oman, Panama, Paraguay, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, and Venezuela.
- Lote
- This field correction involves all models of Silhouette and Sure-T insulin infusion sets.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States