Medfusion¿ 4000 Syringe Infusion Pumps Medfusion¿ 4000 Syringe Infusion Pumps are designed for use in care areas for the administration of fluids requiring precisely controlled infusion rates. These pumps are indicated for use by healthcare professionals in the delivery of IV solutions, drug solutions, blood or blood products, and enteral feedings in continuous, intermittent, and body weight delivery modes.
Publicado el 25 de diciembre de 2013 · Inicio del retiro: 22 de noviembre de 2013
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Smiths Medical is conducting a recall involving all Medfusion¿ 4000 Syringe Infusion Pumps. Smiths Medical has identified an issue with electrical shorting at the AC power receptacle. Shorting can occur when saline solution or other conductive fluids (e.g., potassium chloride) enter the AC power receptacle. When this happens, sparking and/ or flames, accompanied by smoke, may appear at the AC po
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Smiths Medical ASD, Inc.
- Distribuidor
- Worldwide Distribution - USA (nationwide ) including the states of : AZ, NY, TN, MS, AR, WV, VA, VT, PA, DE, DC, OH, CO, NH, MI, CA, FL, CT, WA, MD, NM, RI, MA, TX, NC, IL, KY, WI and MI., and the countries of CANADA, SAUDI ARABIA and AUSTRALIA.
- Lote
- All Medfusion¿ 4000 Syringe Pumps in the field are affected by this Field Safety Corrective Action.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States