FDAFDA · Alimentos y medicamentos·N.º oficial Z-0541-2022

The DIXI Medical Microdeep Depth Electrode (Microdeep Depth Electrode or Electrode) is a smooth electrode that is 0.8 mm in diameter with a rounded (hemispherical) tip. The electrode has multiple platinum/iridium contacts that are 2 mm long, with a total exploration length varying from 16 mm to 80.5 mm according to the electrode reference. Microdeep Depth Electrodes are intended for temporary use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain.

Publicado el 9 de febrero de 2022 · Inicio del retiro: 9 de diciembre de 2021

Resumen

FDA registra el retiro Z-0541-2022 de dispositivos médicos de Clase II, publicado el 9 de febrero de 2022 e iniciado el 9 de diciembre de 2021. Empresa responsable: DIXI MEDICAL USA; Estado: Finalizado; Cantidad reportada: 1792 units; Lote: All lots UDI:03664539000176. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Potential deformation of the internal packaging (blister pack) could compromise sterile barrier of the device when removed from the Tyvek bag

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
DIXI MEDICAL USA
Distribuidor
US Nationwide distribution.
Lote
All lots UDI:03664539000176
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿The DIXI Medical Microdeep Depth Electrode (Microdeep Depth Electrode or Electrode) is a smooth electrode that is 0.8 mm in diameter with a rounded (hemispherical) tip. The electrode has multiple platinum/iridium contacts that are 2 mm long, with a total exploration length varying from 16 mm to 80.5 mm according to the electrode reference. Microdeep Depth Electrodes are intended for temporary use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain. fue retirado del mercado?

Sí. FDA publicó el retiro Z-0541-2022 el 9 de febrero de 2022. Peligro reportado: Potential deformation of the internal packaging (blister pack) could compromise sterile barrier of the device when removed from the Tyvek bag

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter