HeartWare Model 1100
Publicado el 28 de diciembre de 2022 · Inicio del retiro: 29 de noviembre de 2022
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Over time, the driveline cover may become hardened, creating difficulty or an inability to slide the cover back to access the driveline to controller connector. Access to the driveline connector may be delayed, which may result in patient harm from a prolonged pump stop, if an urgent driveline connection or controller issue needs to be addressed. In the event a hardened driveline cover is observed, contact your Medtronic field representative for further assessment and/or servicing.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Heartware, Inc.
- Distribuidor
- Worldwide distribution.
- Lote
- Serial Numbers: HW001, HW012, HW016, HW017, HW023, HW025, HW028, HW044, HW045, HW046, HW047, HW049, HW050, HW051, HW061, HW064, HW083, HW084, HW086, HW091, HW094, HW104, HW105, HW108, HW109, HW110, HW114, HW115, HW116, HW119, HW126, HW127, HW129, HW130, HW131, HW138, HW139, HW143, HW146, HW147, HW153, HW156, HW161, HW170, HW181
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States