Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0760-2020

Airway Gas Option N-CAiO

Publicado el 22 de enero de 2020 · Inicio del retiro: 27 de septiembre de 2019

Resumen

FDA registra el retiro Z-0760-2020 de dispositivos médicos de Clase I, publicado el 22 de enero de 2020 e iniciado el 27 de septiembre de 2019. Empresa responsable: GE Healthcare, LLC; Estado: Finalizado; Cantidad reportada: 12 units; Lote: Mfg. Lot or Serial # UDI # SGV19268009HA 010084068210407421SGV19268009HA11190626 SGV19248001HA 010084068210407421SGV19248001HA11190611 SGV19258009HA 010084068210407421SGV19258009HA11190618 SGV19248004HA 010084068210407421SGV19248004HA11190611 SGV19258006HA 010084068210407421SGV19258006HA11190618 SGV19258003HA 010084068210407421SGV19258003HA11190618 SGV19258005HA 010084068210407421SGV19258005HA11190618 SGV19258012HA 010084068210407421SGV19258012HA11190618 SGV19258020HA 010084068210407421SGV19258020HA11190618 SGV19268003HA 010084068210407421SGV19268003HA11190626 SGV19268004HA 010084068210407421SGV19268004HA11190626 SGV19268005HA 010084068210407421SGV19268005HA11190626. La Clase I indica una probabilidad razonable de consecuencias graves para la salud o muerte.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Displayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could cause up to 50% measurement error in the EtO2/FiO2 values.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
GE Healthcare, LLC
Distribuidor
US OUS: Australia, Austria, Belgium, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Finland, France, Germany, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Korea (Republic of), Malaysia, Mexico, Myanmar, Netherlands, New Zealand, Peru, Poland, Republic of Serbia, Romania, Russia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Uganda, and Unites Kingdom
Lote
Mfg. Lot or Serial # UDI # SGV19268009HA 010084068210407421SGV19268009HA11190626 SGV19248001HA 010084068210407421SGV19248001HA11190611 SGV19258009HA 010084068210407421SGV19258009HA11190618 SGV19248004HA 010084068210407421SGV19248004HA11190611 SGV19258006HA 010084068210407421SGV19258006HA11190618 SGV19258003HA 010084068210407421SGV19258003HA11190618 SGV19258005HA 010084068210407421SGV19258005HA11190618 SGV19258012HA 010084068210407421SGV19258012HA11190618 SGV19258020HA 010084068210407421SGV19258020HA11190618 SGV19268003HA 010084068210407421SGV19268003HA11190626 SGV19268004HA 010084068210407421SGV19268004HA11190626 SGV19268005HA 010084068210407421SGV19268005HA11190626
Clasificación de riesgo
Class I
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Airway Gas Option N-CAiO fue retirado del mercado?

Sí. FDA publicó el retiro Z-0760-2020 el 22 de enero de 2020. Peligro reportado: Displayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could cause up to 50% measurement error in the EtO2/FiO2 values.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter