The PLEX Elite 9000, a Swept-Source OCT [SS-OCT] - Product Usage: The PLEX Elite 9000 Swept-Source OCT [SS-OCT] is a non- contact, high resolution, wide field of view tomographic and biomicroscopic imaging device intended for in-vivo viewing, axial cross-sectional and three-dimensional imaging of posterior ocular structures. The device is indicated for visualizing posterior ocular structures including, but not limited to, retina, retinal nerve fiber layer, ganglion cell plus inner plexiform layer, macula, optic nerve head, vitreous and choroid. The PLEX Elite SS-OCT angiography is indicated as an aid in the visualization of vascular structures of the retina and choroid.
Publicado el 8 de enero de 2020 · Inicio del retiro: 16 de octubre de 2018
Resumen
FDA registra el retiro Z-0770-2020 de dispositivos médicos de Clase II, publicado el 8 de enero de 2020 e iniciado el 16 de octubre de 2018. Empresa responsable: Carl Zeiss Meditec, Inc.; Estado: Finalizado; Cantidad reportada: 151 devices; Lote: OBO /HJO Serial Numbers: PE9000-001, PE9000-002, PE9000-003, PE9000-003,PE9000-004, PE9000-005, PE9000-006, PE9000-007, PE9000-008, PE9000-009, PE9000-0010, PE9000-0011, PE9000-0012, PE9000-0013, PE9000-0014, PE9000-0015, PE9000-0016, PE9000-0020, PE9000-0021, PE9000-0022, PE9000-0023, PE9000-0024, PE9000-0025, PE9000-0026, PE9000-0027, PE9000-0029, PE9000-0030, PE9000-0031, PE9000-0032, PE9000-0033, PE9000-0034, PE9000-0035, PE9000-0036, PE9000-0038, PE9000-0039, PE9000-0040, PE9000-0041, PE9000-0042, PE9000-0043, PE9000-0044, PE9000-0045, PE9000-0046, PE9000-0047, PE9000-0048, PE9000-0049, PE9000-0050, PE9000-0051, PE9000-0052, PE9000-0053, PE9000-0054, PE9000-0055, PE9000-0056, PE9000-0057, PE9000-0058, PE9000-0059, PE9000-0060, PE9000-0061, PE9000-0062, PE9000-0063, PE9000-0064, PE9000-0065, PE9000-0066, PE9000-0067, PE9000-0068, PE9000-0069, PE9000-0070, PE9000-0071, PE9000-0072, PE9000-0073, PE9000-0074, PE9000-0075, PE9000-0076, PE9000-0077, PE9000-0078, PE9000-0079, PE9000-0080, PE9000-0081, PE9000-0082, PE9000-0083, PE9000-0084, PE9000-0085, PE9000-0086, PE9000-0087, PE9000-0088, PE9000-0089, PE9000-0090, PE9000-0091, PE9000-0092, PE9000-0093, PE9000-0094, PE9000-0095, PE9000-0096, PE9000-0097, PE9000-0098, PE9000-0099, PE9000-0100, PE9000-0101, PE9000-0102, PE9000-0103, PE9000-0104, PE9000-0105, PE9000-0106, PE9000-0107, PE9000-0108, PE9000-0109, PE9000-0110, PE9000-0111, PE9000-0112, PE9000-0113, PE9000-0114, PE9000-0115, PE9000-0116, PE9000-0117, PE9000-0118, PE9000-0119, PE9000-0120, PE9000-0121, PE9000-0122, PE9000-0123, PE9000-0124, PE9000-0125, PE9000-0126, PE9000-0127, PE9000-0128, PE9000-0129, PE9000-0130, PE9000-0131, PE9000-0132, PE9000-0133, PE9000-0134, PE9000-0135, PE9000-0137, PE9000-0138, PE9000-0139, PE9000-0140, PE9000-0141, PE9000-0142, PE9000-0143, PE9000-0144, PE9000-0145, PE9000-0146, PE9000-0147, PE9000-0148, PE9000-0149, PE9000-0150, PE9000-0151, PE9000-0152, PE9000-0153, PE9000-0154, PE9000-0155, PE9000-0156, PE9000-0158. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Under certain fault conditions, the existing design may expose the operator to a hazardous voltage when contacting the metal enclosure (including mounting screws) around the up/down switch.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Carl Zeiss Meditec, Inc.
- Distribuidor
- Worldwide distribution - US Nationwide distribution in the states of CA, TX, NY, MA, OH, PA, IA, , KY, FL, MI, IL, MD and Germany, Japan, Switzerland, France Italy, Australia, Austria, Korea, China, India, Canada, Singapore, Norway, Taiwan, Israel, UAE.
- Lote
- OBO /HJO Serial Numbers: PE9000-001, PE9000-002, PE9000-003, PE9000-003,PE9000-004, PE9000-005, PE9000-006, PE9000-007, PE9000-008, PE9000-009, PE9000-0010, PE9000-0011, PE9000-0012, PE9000-0013, PE9000-0014, PE9000-0015, PE9000-0016, PE9000-0020, PE9000-0021, PE9000-0022, PE9000-0023, PE9000-0024, PE9000-0025, PE9000-0026, PE9000-0027, PE9000-0029, PE9000-0030, PE9000-0031, PE9000-0032, PE9000-0033, PE9000-0034, PE9000-0035, PE9000-0036, PE9000-0038, PE9000-0039, PE9000-0040, PE9000-0041, PE9000-0042, PE9000-0043, PE9000-0044, PE9000-0045, PE9000-0046, PE9000-0047, PE9000-0048, PE9000-0049, PE9000-0050, PE9000-0051, PE9000-0052, PE9000-0053, PE9000-0054, PE9000-0055, PE9000-0056, PE9000-0057, PE9000-0058, PE9000-0059, PE9000-0060, PE9000-0061, PE9000-0062, PE9000-0063, PE9000-0064, PE9000-0065, PE9000-0066, PE9000-0067, PE9000-0068, PE9000-0069, PE9000-0070, PE9000-0071, PE9000-0072, PE9000-0073, PE9000-0074, PE9000-0075, PE9000-0076, PE9000-0077, PE9000-0078, PE9000-0079, PE9000-0080, PE9000-0081, PE9000-0082, PE9000-0083, PE9000-0084, PE9000-0085, PE9000-0086, PE9000-0087, PE9000-0088, PE9000-0089, PE9000-0090, PE9000-0091, PE9000-0092, PE9000-0093, PE9000-0094, PE9000-0095, PE9000-0096, PE9000-0097, PE9000-0098, PE9000-0099, PE9000-0100, PE9000-0101, PE9000-0102, PE9000-0103, PE9000-0104, PE9000-0105, PE9000-0106, PE9000-0107, PE9000-0108, PE9000-0109, PE9000-0110, PE9000-0111, PE9000-0112, PE9000-0113, PE9000-0114, PE9000-0115, PE9000-0116, PE9000-0117, PE9000-0118, PE9000-0119, PE9000-0120, PE9000-0121, PE9000-0122, PE9000-0123, PE9000-0124, PE9000-0125, PE9000-0126, PE9000-0127, PE9000-0128, PE9000-0129, PE9000-0130, PE9000-0131, PE9000-0132, PE9000-0133, PE9000-0134, PE9000-0135, PE9000-0137, PE9000-0138, PE9000-0139, PE9000-0140, PE9000-0141, PE9000-0142, PE9000-0143, PE9000-0144, PE9000-0145, PE9000-0146, PE9000-0147, PE9000-0148, PE9000-0149, PE9000-0150, PE9000-0151, PE9000-0152, PE9000-0153, PE9000-0154, PE9000-0155, PE9000-0156, PE9000-0158
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDABlisovi Fe 1.5/30 (norethindrone acetate and ethinyl estradiol tablets USP and ferrous fumarate tablets 75mg),15 de enero de 2020
- FDAJuno DRF, Model Number 709020 - Product Usage: General. multi-function x-ray systems capable of routine8 de enero de 2020
- FDAZeiss brand IOLMaster 500, Model 500, Biomicroscope, Slit-Lamp, AC-Powered, Product is manufactured by Carl27 de noviembre de 2013
- FDAZeiss brand IOLMaster, Version 1-5, Biomicroscope, Slit-Lamp, AC-Powered, Product is manufactured by Carl27 de noviembre de 2013
- FDAZeiss brand OPMI Pentero C, Surgical microscope and accessories, Model: OPMI Pentero C, Product Numbers:12 de junio de 2013
- FDAINTRABEAM Balloon Applicator Set; 3.5 cm, Carl Zeiss Meditec AG, Oberkochen, Germany INTRABEAM System is28 de noviembre de 2012
Preguntas sobre este retiro
¿The PLEX Elite 9000, a Swept-Source OCT [SS-OCT] - Product Usage: The PLEX Elite 9000 Swept-Source OCT [SS-OCT] is a non- contact, high resolution, wide field of view tomographic and biomicroscopic imaging device intended for in-vivo viewing, axial cross-sectional and three-dimensional imaging of posterior ocular structures. The device is indicated for visualizing posterior ocular structures including, but not limited to, retina, retinal nerve fiber layer, ganglion cell plus inner plexiform layer, macula, optic nerve head, vitreous and choroid. The PLEX Elite SS-OCT angiography is indicated as an aid in the visualization of vascular structures of the retina and choroid. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0770-2020 el 8 de enero de 2020. Peligro reportado: Under certain fault conditions, the existing design may expose the operator to a hazardous voltage when contacting the metal enclosure (including mounting screws) around the up/down switch.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter