Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0895-2013

Orthofix ISKD Limb Lengthener. ISKD - Intramedullary Skeletal Kinetic Distractor. The ISKD is intended for limb lengthening of the tibia and femur.

Publicado el 6 de marzo de 2013 · Inicio del retiro: 6 de agosto de 2012

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

There is a possibility that the ISKD limb lengthener may stop distracting post-operatively during treatment, which may result in premature bone consolidation (limb not achieving the desired length) leading to revision surgery to remove and/or replace the device.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Orthofix, Inc
Distribuidor
Worldwide Distribution - USA (nationwide) Internationally to Australia, Hong Kong, Saudi Arabia, Korea, Turkey, Greece, Finland, South Africa, Spain, Singapore, Switzerland, Brazil, Germany, France, Japan, Netherlands, New Zealand, Israel, Canada and Mexico
Lote
All serial numbers of the following Catalog Numbers: F12-255-305, F12-345-395NS, T12-245-295, F12-255-305NS, F12-345-425, T12-255-305, F12-255-335, F12-345-425NS, T12-255-305NS, F12-255-335NS, T10-215-265, T12-255-335, F12-300-350, T10-255-305, T12-255-335NS, F12-300-350NS, T10-255-335, T12-300-350, F12-300-380, T10-300-350, T12-300-350NS, F12-300-380NS, T12-215-265, T12-300-380, F12-345-395, T12-215-265NS, and T12-300-380NS
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States