CARESCAPE Monitor B850, a multi-parameter high acuity patient monitor intended for use in multiple areas within a professional healthcare facility. Product Usage: The CARESCAPE Monitor B850 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time. The CARESCAPE Monitor B850 system is indicated for monitoring of Hemodynamic (including ECG, ST Segment, Arrhythmia Detection, ECG Diagnostic Analysis and Measurement, Invasive Pressure, Noninvasive Blood Pressure, Pulse Oximetry, Cardiac Output, Temperature, Impedance Respiration and SvO2 (Mixed Venous Oxygen Saturation)), Airway Gases (Fi/Et C02, 02, N20 and Anesthetic Agent), Spirometry, Gas Exchange (02 Consumption (VO2), C02 production (VCO2), energy expenditure (BE), and respiratory quotient (RQ)) and neurophysiological (including electroencephalography (EEG), Entropy, Bispectral Index (BIS) and Neuromuscular Transmission (NMT) Monitoring) status. The CARESCAPE Monitor B850 provides alarms, trends, snapshots and events, and calculations and can be connected to displays, printers and recording devices. The CARESCAPE Monitor B850 can be a stand-alone monitor or interfaced to other devices. It can also be connected to other monitors for bed to bed viewing and to data management software devices via a network. The CARESCAPE Monitor B850 is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility. In addition to the healthcare practitioner, the CARESCAPE Monitor B850 is designed to provide configuration and troubleshooting capabilities to qualified service personnel.
Publicado el 12 de febrero de 2014 · Inicio del retiro: 20 de diciembre de 2013
Resumen
FDA registra el retiro Z-0927-2014 de dispositivos médicos de Clase II, publicado el 12 de febrero de 2014 e iniciado el 20 de diciembre de 2013. Empresa responsable: GE Healthcare, LLC; Estado: Finalizado; Cantidad reportada: 12, 945 units (6,239 US, 6,706 OUS); Lote: serial numbers: SEG09270517HX, SEG09270549HX, SEG09270551HX, SEG09270553HX, SEG09270554HX, SEG09270557HX, SEG09270559HX, SEG09270561HX, SEG09270562HX, SEG09270563HX, SEG09270564HX, SEG09270565HX, SEG09270566HX, SEG09270567HX, SEG09270568HX, SEG09270569HX, SEG09270583HX, SEG09270584HX, SEG09270585HX, SEG09270587HX, SEG09270588HX, SEG09270589HX, SEG09270591HX, SEG09270592HX, SEG09280596HX, SEG09280597HX, SEG09280598HX, SEG09280599HX, SEG09280600HX, SEG09280601HX, SEG09280602HX, SEG09280603HX, SEG09280604HX, SEG09280605HX, SEG09280606HX, SEG09280607HX, SEG09280608HX, SEG09280609HX, SEG09280610HX, SEG09280611HX, SEG09280612HX, SEG09280613HX, SEG09280614HX, SEG09280615HX, SEG09280616HX, SEG09280617HX, SEG09280618HX, SEG09280619HX, SEG09280620HX, SEG09280621HX, SEG09280622HX, SEG09280623HX, SEG09280624HX, SEG09280625HX, SEG09310656HX, SEG09310657HX, SEG09310658HX, SEG09310659HX, SEG09310660HX, SEG09310661HX, SEG09310662HX, SEG09310663HX, SEG09310664HX, SEG09310665HX, SEG09310666HX, SEG09310667HX, SEG09310668HX, SEG09310669HX, SEG09310670HX, SEG09310671HX, SEG09310672HX, SEG09310673HX, SEG09310674HX, SEG09310675HX, SEG09310696HX, SEG09310697HX, SEG09310698HX, SEG09310699HX, SEG09310700HX, SEG09310701HX, SEG09310702HX, SEG09310703HX, SEG09310704HX, SEG09310705HX, SEG09310706HX, SEG09310707HX, SEG09310708HX, SEG09310709HX, SEG09310710HX, SEG09310711HX, SEG09310712HX, SEG09310713HX, SEG09310714HX, SEG09310715HX, SEG09310716HX, SEG09310717HX, SEG09310718HX, SEG09310719HX, SEG09310720HX, SEG09310721HX, SEG09310722HX, SEG09310723HX, SEG09310724HX, SEG09310725HX, SEG09310726HX, SEG09310727HX, SEG09310728HX, SEG09310729HX, SEG09310730HX, SEG09310731HX, SEG09310732HX, SEG09310733HX, SEG09310734HX, SEG09310735HX, SEG09310736HX, SEG09310737HX, SEG09310738HX, SEG09310739HX, SEG09310740HX, SEG09310741HX, SEG09310742HX, SEG09310743HX, SEG09310744HX, SEG09310745HX, SEG09320797HX, SEG09320798HX, SEG09320799HX, SEG09320800HX, SEG09320801HX, SEG09320802HX, SEG09320803HX, SEG09320804HX, SEG09320805HX, SEG09320806HX, SEG09320807HX, SEG09320808HX, SEG09320809HX, SEG09320810HX, SEG09320811HX, SEG09320812HX, SEG09320813HX, SEG09320814HX, SEG09320815HX, SEG09320816HX, SEG09320817HX, SEG09320818HX, SEG09320819HX, SEG09320820HX, SEG09320821HX, SEG09320822HX, SEG09320823HX, SEG09320824HX, SEG09320825HX, SEG09320826HX, SEG09320827HX, SEG09320828HX, SEG09320829HX, SEG09320830HX, SEG09320831HX, SEG09320832HX, SEG09320833HX, SEG09320834HX, SEG09320835HX, SEG09320836HX, SEG09320837HX, SEG09320838HX, SEG09320839HX, SEG09320840HX, SEG09320841HX, SEG09320842HX, SEG09320843HX, SEG09320844HX, SEG09320845HX, SEG09320846HX, SEG09320847HX, SEG09320848HX, SEG09320849HX, SEG09320850HX, SEG09320851HX, SEG09320852HX, SEG09320853HX, SEG09320854HX, SEG09320855HX, SEG09320856HX, SEG09320857HX, SEG09320858HX, SEG09320859HX, SEG09320860HX, SEG09320861HX, SEG09320862HX, SEG09320863HX, SEG09320864HX, SEG09320865HX, SEG09320866HX, SEG09320867HX, SEG09320868HX, SEG09320869HX, SEG09320870HX, SEG09320871HX, SEG09320872HX, SEG09320873HX, SEG09320874HX, SEG09320875HX, SEG09320876HX, SEG09320877HX, SEG09320878HX, SEG09320879HX, SEG09320880HX, SEG09320881HX, SEG09320882HX, SEG09320883HX, SEG09320884HX, SEG09320885HX, SEG09320886HX, SEG09320887HX, SEG09320888HX, SEG09320889HX, SEG09320890HX, SEG09320891HX, SEG09320892HX, SEG09320893HX, SEG09320894HX, SEG09320895HX, SEG09320896HX, SEG09360898HX, SEG09360899HX, SEG09360900HX, SEG09360901HX, SEG09360902HX, SEG09360903HX, SEG09360904HX, SEG09360905HX, SEG09360906HX, SEG09360907HX, SEG09360908HX, SEG09360909HX, SEG09360910HX, SEG09360911HX, SEG09360912HX, SEG09360913HX, SEG09360914HX, SEG09360915HX, SEG09360916HX, SEG09360917HX, SEG09360918HX, SEG09360919HX, SEG09360920HX, SEG09360921HX, SEG09360922HX, SEG09360923HX, SEG09360924HX, SEG09360925HX, SEG09360926HX, SEG09360927HX, SEG09360928HX, SEG09360929HX, SEG09360930HX, SEG09360931HX, SEG09360932HX, SEG09360933HX, SEG09360934HX, SEG09360935HX, SEG09360936HX, SEG09360937HX, SEG09360938HX, SEG09360. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
There is a potential for communication loss associated with the F5-01 Frame when connected to CARESCAPE Patient Data Module (PDM) in the CARESCAPE Monitor B850. The PDM patient parameter acquisition module can sometimes spontaneously lose the communication with the host patient monitor when attached to the F5-01 Frame due to an issue with the PDM Ethernet communication hardware. The communicatio
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- GE Healthcare, LLC
- Distribuidor
- Worldwide Distribution - US Nationwide including Puerto Rico and the countries of: Australia, Austria, Azerbaijan, Bahrain, Belgium, BRAZIL, Canada, CHILE, COSTA RICA, Cyprus, Czech Republic, Denmark, ECUADOR, Egypt, Estonia, Finland, France, Germany, Greece, GUAM, India, INDONESIA, Iraq, Ireland, Israel, Italy, Japan, Kazakhstan, Korea, Republic of, Kuwait, Lithuania, Malaysia, Malta, Mexico, NEPAL, Netherlands, New Zealand, Norway, Pakistan, PANAMA, PARAGUAY, PERU, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tanzania, United Republic of, Turkey, United Arab Emirates, United Kingdom, URUGUAY.
- Lote
- serial numbers: SEG09270517HX, SEG09270549HX, SEG09270551HX, SEG09270553HX, SEG09270554HX, SEG09270557HX, SEG09270559HX, SEG09270561HX, SEG09270562HX, SEG09270563HX, SEG09270564HX, SEG09270565HX, SEG09270566HX, SEG09270567HX, SEG09270568HX, SEG09270569HX, SEG09270583HX, SEG09270584HX, SEG09270585HX, SEG09270587HX, SEG09270588HX, SEG09270589HX, SEG09270591HX, SEG09270592HX, SEG09280596HX, SEG09280597HX, SEG09280598HX, SEG09280599HX, SEG09280600HX, SEG09280601HX, SEG09280602HX, SEG09280603HX, SEG09280604HX, SEG09280605HX, SEG09280606HX, SEG09280607HX, SEG09280608HX, SEG09280609HX, SEG09280610HX, SEG09280611HX, SEG09280612HX, SEG09280613HX, SEG09280614HX, SEG09280615HX, SEG09280616HX, SEG09280617HX, SEG09280618HX, SEG09280619HX, SEG09280620HX, SEG09280621HX, SEG09280622HX, SEG09280623HX, SEG09280624HX, SEG09280625HX, SEG09310656HX, SEG09310657HX, SEG09310658HX, SEG09310659HX, SEG09310660HX, SEG09310661HX, SEG09310662HX, SEG09310663HX, SEG09310664HX, SEG09310665HX, SEG09310666HX, SEG09310667HX, SEG09310668HX, SEG09310669HX, SEG09310670HX, SEG09310671HX, SEG09310672HX, SEG09310673HX, SEG09310674HX, SEG09310675HX, SEG09310696HX, SEG09310697HX, SEG09310698HX, SEG09310699HX, SEG09310700HX, SEG09310701HX, SEG09310702HX, SEG09310703HX, SEG09310704HX, SEG09310705HX, SEG09310706HX, SEG09310707HX, SEG09310708HX, SEG09310709HX, SEG09310710HX, SEG09310711HX, SEG09310712HX, SEG09310713HX, SEG09310714HX, SEG09310715HX, SEG09310716HX, SEG09310717HX, SEG09310718HX, SEG09310719HX, SEG09310720HX, SEG09310721HX, SEG09310722HX, SEG09310723HX, SEG09310724HX, SEG09310725HX, SEG09310726HX, SEG09310727HX, SEG09310728HX, SEG09310729HX, SEG09310730HX, SEG09310731HX, SEG09310732HX, SEG09310733HX, SEG09310734HX, SEG09310735HX, SEG09310736HX, SEG09310737HX, SEG09310738HX, SEG09310739HX, SEG09310740HX, SEG09310741HX, SEG09310742HX, SEG09310743HX, SEG09310744HX, SEG09310745HX, SEG09320797HX, SEG09320798HX, SEG09320799HX, SEG09320800HX, SEG09320801HX, SEG09320802HX, SEG09320803HX, SEG09320804HX, SEG09320805HX, SEG09320806HX, SEG09320807HX, SEG09320808HX, SEG09320809HX, SEG09320810HX, SEG09320811HX, SEG09320812HX, SEG09320813HX, SEG09320814HX, SEG09320815HX, SEG09320816HX, SEG09320817HX, SEG09320818HX, SEG09320819HX, SEG09320820HX, SEG09320821HX, SEG09320822HX, SEG09320823HX, SEG09320824HX, SEG09320825HX, SEG09320826HX, SEG09320827HX, SEG09320828HX, SEG09320829HX, SEG09320830HX, SEG09320831HX, SEG09320832HX, SEG09320833HX, SEG09320834HX, SEG09320835HX, SEG09320836HX, SEG09320837HX, SEG09320838HX, SEG09320839HX, SEG09320840HX, SEG09320841HX, SEG09320842HX, SEG09320843HX, SEG09320844HX, SEG09320845HX, SEG09320846HX, SEG09320847HX, SEG09320848HX, SEG09320849HX, SEG09320850HX, SEG09320851HX, SEG09320852HX, SEG09320853HX, SEG09320854HX, SEG09320855HX, SEG09320856HX, SEG09320857HX, SEG09320858HX, SEG09320859HX, SEG09320860HX, SEG09320861HX, SEG09320862HX, SEG09320863HX, SEG09320864HX, SEG09320865HX, SEG09320866HX, SEG09320867HX, SEG09320868HX, SEG09320869HX, SEG09320870HX, SEG09320871HX, SEG09320872HX, SEG09320873HX, SEG09320874HX, SEG09320875HX, SEG09320876HX, SEG09320877HX, SEG09320878HX, SEG09320879HX, SEG09320880HX, SEG09320881HX, SEG09320882HX, SEG09320883HX, SEG09320884HX, SEG09320885HX, SEG09320886HX, SEG09320887HX, SEG09320888HX, SEG09320889HX, SEG09320890HX, SEG09320891HX, SEG09320892HX, SEG09320893HX, SEG09320894HX, SEG09320895HX, SEG09320896HX, SEG09360898HX, SEG09360899HX, SEG09360900HX, SEG09360901HX, SEG09360902HX, SEG09360903HX, SEG09360904HX, SEG09360905HX, SEG09360906HX, SEG09360907HX, SEG09360908HX, SEG09360909HX, SEG09360910HX, SEG09360911HX, SEG09360912HX, SEG09360913HX, SEG09360914HX, SEG09360915HX, SEG09360916HX, SEG09360917HX, SEG09360918HX, SEG09360919HX, SEG09360920HX, SEG09360921HX, SEG09360922HX, SEG09360923HX, SEG09360924HX, SEG09360925HX, SEG09360926HX, SEG09360927HX, SEG09360928HX, SEG09360929HX, SEG09360930HX, SEG09360931HX, SEG09360932HX, SEG09360933HX, SEG09360934HX, SEG09360935HX, SEG09360936HX, SEG09360937HX, SEG09360938HX, SEG09360
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDACentricity PACS Software Version 7.0 SP0.0.4.73 de agosto de 2022
- FDASIGNA Premier magnetic resonance scanner, model 5748519.27 de julio de 2022
- FDAGE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical images and13 de julio de 2022
- FDABack up batteries in Avance CS2 and Avance CS2 Pro anesthesia devices manufactured on or after April 1, 20196 de julio de 2022
- FDAThe Rotoprone therapy System is an advanced patient care system for the treatment and prevention of pulmonary12 de febrero de 2014
- FDAXiO RTP System. Used to create treatment plans for any cancer patient for whom external beam radiation therapy12 de febrero de 2014
Preguntas sobre este retiro
¿CARESCAPE Monitor B850, a multi-parameter high acuity patient monitor intended for use in multiple areas within a professional healthcare facility. Product Usage: The CARESCAPE Monitor B850 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time. The CARESCAPE Monitor B850 system is indicated for monitoring of Hemodynamic (including ECG, ST Segment, Arrhythmia Detection, ECG Diagnostic Analysis and Measurement, Invasive Pressure, Noninvasive Blood Pressure, Pulse Oximetry, Cardiac Output, Temperature, Impedance Respiration and SvO2 (Mixed Venous Oxygen Saturation)), Airway Gases (Fi/Et C02, 02, N20 and Anesthetic Agent), Spirometry, Gas Exchange (02 Consumption (VO2), C02 production (VCO2), energy expenditure (BE), and respiratory quotient (RQ)) and neurophysiological (including electroencephalography (EEG), Entropy, Bispectral Index (BIS) and Neuromuscular Transmission (NMT) Monitoring) status. The CARESCAPE Monitor B850 provides alarms, trends, snapshots and events, and calculations and can be connected to displays, printers and recording devices. The CARESCAPE Monitor B850 can be a stand-alone monitor or interfaced to other devices. It can also be connected to other monitors for bed to bed viewing and to data management software devices via a network. The CARESCAPE Monitor B850 is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility. In addition to the healthcare practitioner, the CARESCAPE Monitor B850 is designed to provide configuration and troubleshooting capabilities to qualified service personnel. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0927-2014 el 12 de febrero de 2014. Peligro reportado: There is a potential for communication loss associated with the F5-01 Frame when connected to CARESCAPE Patient Data Module (PDM) in the CARESCAPE Monitor B850. The PDM patient parameter acquisition module can sometimes spontaneously lose the communication with the host patient monitor when attached to the F5-01 Frame due to an issue with the PDM Ethernet communication hardware. The communicatio
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter