Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0955-2020

Medtronic MiniMed 630G System with SmartGuard, Ref/Model # MMT-1515, MMT-1715, MMT-1714, MMT-1754, and MMT-1755 *US and Canada Version Only

Publicado el 19 de febrero de 2020 · Inicio del retiro: 21 de noviembre de 2019

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

There have been reported incidents of a loose reservoir that can no longer be locked into the pump. The reservoir can become loose due to a broken or missing retainer ring that prevents a proper lock. If the reservoir is not properly locked into the pump, it could lead to over or under delivery of insulin, which could then result in hypoglycemia or hyperglycemia.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Medtronic Inc.
Distribuidor
US:AK,AL,AR,AZ,BM,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,ND,NE,NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY OUS: France, Germany, UK, Italy, Netherlands, Spain, Sweden, Israel, Norway, Austria, Belgium, Denmark, Czech Republic, Finland, Switzerland, Slovakia, Ireland, Poland, S. Africa, Greece, Hungary, Saudi Arabia, United Arab Emira, New Zealand, Australia, Japan, Mexico, Colombia, Brazil, S. East Asia, Korea, Canada,
Lote
All Serial Numbers ( excluding devices with black retainer rings)
Clasificación de riesgo
Class I
Área (FDA)
device
País
United States