ExcelsiusGPS Cannula A, 9mm x 250mm, Cannula used with ball tip probes, Model No. 6143.2844
Publicado el 14 de febrero de 2024 · Inicio del retiro: 15 de diciembre de 2023
Resumen
FDA registra el retiro Z-1007-2024 de dispositivos médicos de Clase II, publicado el 14 de febrero de 2024 e iniciado el 15 de diciembre de 2023. Empresa responsable: Globus Medical, Inc.; Estado: En curso; Cantidad reportada: 570; Lote: UDI-DI 00889095998498 Lots FSB136AA FSB121AA FSB241AA FSB207AA FSB136AA FSB121AA FSB110AA FSB073AA FSA286CA FSA286BA FSA286AA FSA263DA FSA263EA FSA263CA FSA263AA FSA263BA FSA160AA FSA160BA FSA105BA FSA105AA FSA053AA FSY319AA FSY279BA FSY279AA FSY272AA FSY230AA FSY098AA FSY098CA FSY098BA FSY048AA FSX093AA FSX205AA FSX141BA FSX141CA FSX141AA FSX093AA FSX063AA FSX041AA FSX041BA FSW142AA FSW094AA. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Devices may not be rendered sterile due to an internal gap that may be present on some devices causing increased difficulty cleaning prior to steam sterilization. There is an increased risk for infection, inflammation, local toxicity, systemic toxicity, sensitivity, and cross-contamination.
Qué hacer
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Globus Medical, Inc.
- Distribuidor
- Worldwide - US Nationwide and the countries of Australia, Austria, Belgium, Czech Republic, Germany, India, Indonesia, Ireland, Italy, Japan, Malaysia, Netherlands, Poland, Saudi Arabia, Singapore, Spain, United Arab Emirates, United Kingdom.
- Lote
- UDI-DI 00889095998498 Lots FSB136AA FSB121AA FSB241AA FSB207AA FSB136AA FSB121AA FSB110AA FSB073AA FSA286CA FSA286BA FSA286AA FSA263DA FSA263EA FSA263CA FSA263AA FSA263BA FSA160AA FSA160BA FSA105BA FSA105AA FSA053AA FSY319AA FSY279BA FSY279AA FSY272AA FSY230AA FSY098AA FSY098CA FSY098BA FSY048AA FSX093AA FSX205AA FSX141BA FSX141CA FSX141AA FSX093AA FSX063AA FSX041AA FSX041BA FSW142AA FSW094AA
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAGlobus Medical Cranial Drill, 6.00mm - instruments for access, preparation, and placement of cranial devices14 de febrero de 2024
- FDAGlobus Medical Cranial Drill with Stop, 3.20mm - instruments for access, preparation, and placement of cranial14 de febrero de 2024
- FDAGlobus Medical Cranial Drill with Stop, 2.40mm - Instruments for access, preparation, and placement of cranial14 de febrero de 2024
- FDAArticle No. NA6050411, Valiant PhD No. 1 Econo14 de febrero de 2024
- FDABrasseler 8S RA SHP RD (Super-Sharp Round) H1S.21.023, REF 001156T0), carbide dental bur, 10 burs/package, Rx.14 de febrero de 2024
- FDAExcelsiusGPS Flat Panel Fluoroscopy Fixture Kit, Part number 6203.210024 de enero de 2024
Preguntas sobre este retiro
¿ExcelsiusGPS Cannula A, 9mm x 250mm, Cannula used with ball tip probes, Model No. 6143.2844 fue retirado del mercado?
Sí. FDA publicó el retiro Z-1007-2024 el 14 de febrero de 2024. Peligro reportado: Devices may not be rendered sterile due to an internal gap that may be present on some devices causing increased difficulty cleaning prior to steam sterilization. There is an increased risk for infection, inflammation, local toxicity, systemic toxicity, sensitivity, and cross-contamination.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit