Chord-X Chordal Sizer is designed to assist in determining the proper length of chordal replacement prostheses to repair the mitral valve and help the surgeon select a preconfigured prosthesis, or configure the correct size using straight suture material, during mitral valve repair.
Publicado el 1 de febrero de 2017 · Inicio del retiro: 13 de diciembre de 2016
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Four issues were uncovered by the recalling firm's CAPA: 1) Burrs were found to be present on Left Housing, Right Housing and Barrel components, 2) Loose plastic particulate was found within sealed primary packages, 3) Loose hair was discovered within four sealed primary packages and 4) Assembly error, wherein Left and Right Housing components were acceptably press-fit together, was discovered within one sealed primary package. One of the Left Housing press-fit pins was not pressed into the associated hole of the Right Housing, resulting in a small gap in the housing assembly.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- On-X Life Technologies, Inc.
- Distribuidor
- CA, WA, NE, DE, OH, ME, TX
- Lote
- 72-00-47420
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States