Intellispace Portal (ISP) Philips Medical Systems (Cleveland), Inc., IntelliSpace Portal (ISP) is a multimodality (CT, NM, and MR) thin-client applications server that delivers full diagnostic viewing and clinical applications to the enterprise. IntelliSpace Portal (ISP) is a medical software system that allows multiple users to remotely access IntelliSpace Workspace Portal (ISP) from compatible computers on a network. The system allows networking, selection, processing and filming of multimodality DICOM images. Both the client and server software are only for use with off the shelf hardware technology that meets defined minimum specifications. trained physicians may use the images as a basis for diagnosis upon ensuring that monitor quality, ambient light conditions and image compression ratios are consistent with clinical application.
Publicado el 24 de abril de 2013 · Inicio del retiro: 12 de febrero de 2013
Resumen
FDA registra el retiro Z-1102-2013 de dispositivos médicos de Clase II, publicado el 24 de abril de 2013 e iniciado el 12 de febrero de 2013. Empresa responsable: Philips Medical Systems (Cleveland) Inc; Estado: Finalizado; Cantidad reportada: 313 units; Lote: Model #881001, Serial #s: 79916, 79932, 79933, 79935, 79936, 79937, 79939, 79940, 79943, 80071, 80914, 80916, 80925, 80926, 80928, 80932, 80939, 80963, 80965, 80977, 80979, 80994, 80995, 81004, 81006, 81008, 81010, 81026, 81027, 81028, 81034, 81035, 81036, 81039, 81063, 81065, 81076, 81079, 81081, 81085, 81089, 81097, 81098, 81108, 81110, 81117, 81119, 81127, 81130, 81132, 81133, 81150, 81153, 81154, 81160, 81161, 81163, 81165, 81169, 81172, 81173, 81174, 81175, 81178, 81179, 81182, 81183, 81184, 81187, 81189, 81190, 81192, 81194, 81196, 81200, 82001, 82002, 82003, 82004, 82006, 82007, 82008, 82009, 82010, 82011, 82012, 82013, 82014, 82015, 82016, 82017, 82018, 82019, 82020, 82021, 82022, 82023, 82024, 82025, 82026, 82027, 82028, 82029, 82030, 82031, 82032, 82033, 82034, 82035, 82036, 82037, 82038, 82050, 82051, 82052, 82053, 82054, 82055, 82056, 82057, 82058, 82059, 82060, 82061, 82062, 82063, 82064, 82065, 82066, 82067, 82068, 82069, 82070, 82071, 82072, 82073, 82074, 82084, 82086, 82087, 82088, 82089, 82090, 82091, 82092, 82095, 82097, 82098, 82099, 82100, 82101, 82102, 82103, 82104, 82105, 82106, 82107, 82108, 82109, 82110, 82112, 82116, 82118, 82120, 82121, 82123, 82125, 82126, 82128, 82129, 82130, 82131, 82132, 82133, 82138, 82142, 82143, 82144, 82146, 82148, 82201, 82202, 82203, 82236, 82238, 82240, 82241, 82250, 82251, 85005, 85006, 85007, 85008, 85009, 85010, 85011, 85014, 85025, 85026, 85027, 85028, 85029, 85031, 85032, 85033, 85040, 85041, 85042, 85043, 85044, 85046, 85047, 85049, 85050, 85066, 85067, 85068, 85069, 85080, 85081, 85082, 85083, 85084, 85086, 85088, 85090, 85096, 85103, 85104, 85105, 85106, 85107, 85108, 85110, 85117, 85141, 85143, 85144, 85145, 85149, 85164, 85165, 85167, 85168, 85169, 85170, 85171, 85172, 85174, 85175, 85176, 85177, 85178, 85179, 85180, 85213, 85214, 85215, 85216, 85217, 85218, 85219, 85238, 85239, 85243, 85244, 85245, 85246, 85247, 85248, 85249, 85250, 85252, 85253, 85254, 85255, 85257, 85260, 85296, 85299, 85300, 85302, 85305, 85306, 85307, 85308, 85309, 85312, 85315, 85316, 85317, 85319, 85320, 85322, 85329, 85330, 85332, 85333, 85334, 85335, 85337, 85338, 85355, 85373, 85475, 95001, 95007, 95008, 95009, 820001, 822002, 822003, 822004, 822005, 822009, 822017, 850004, 850015, 920003 & 950001. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Philips was notified of a problem using the IntelliSpace Portal with software version 4.0.2 with the URL PACS integration. Philips was informed that when loading a study to an analysis application while using URL PACS integration configuration, after another study was opened by a review application, the Analysis monitor is updated with the study of the current patient, but the Review monitor conti
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Philips Medical Systems (Cleveland) Inc
- Distribuidor
- Worldwide Distribution - (USA) Nationwide Distribution including the states of AL, AR, AZ, CA, FL, GA, IA, IN, LA, MI, MN, MO, NC, NJ, NY, OH, OK, OR, PA, TN, TX, WA, WI & WV. and the countries of Australia, Austria, Belgium, Canada, China, Czech Republic, Denmark, Egypt, Finland, France, France, Germany, Greece, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Latvia, Lebanon, Monaco, Netherlands, Norway, Poland, Portugal, Russia, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand & United Kingdom
- Lote
- Model #881001, Serial #s: 79916, 79932, 79933, 79935, 79936, 79937, 79939, 79940, 79943, 80071, 80914, 80916, 80925, 80926, 80928, 80932, 80939, 80963, 80965, 80977, 80979, 80994, 80995, 81004, 81006, 81008, 81010, 81026, 81027, 81028, 81034, 81035, 81036, 81039, 81063, 81065, 81076, 81079, 81081, 81085, 81089, 81097, 81098, 81108, 81110, 81117, 81119, 81127, 81130, 81132, 81133, 81150, 81153, 81154, 81160, 81161, 81163, 81165, 81169, 81172, 81173, 81174, 81175, 81178, 81179, 81182, 81183, 81184, 81187, 81189, 81190, 81192, 81194, 81196, 81200, 82001, 82002, 82003, 82004, 82006, 82007, 82008, 82009, 82010, 82011, 82012, 82013, 82014, 82015, 82016, 82017, 82018, 82019, 82020, 82021, 82022, 82023, 82024, 82025, 82026, 82027, 82028, 82029, 82030, 82031, 82032, 82033, 82034, 82035, 82036, 82037, 82038, 82050, 82051, 82052, 82053, 82054, 82055, 82056, 82057, 82058, 82059, 82060, 82061, 82062, 82063, 82064, 82065, 82066, 82067, 82068, 82069, 82070, 82071, 82072, 82073, 82074, 82084, 82086, 82087, 82088, 82089, 82090, 82091, 82092, 82095, 82097, 82098, 82099, 82100, 82101, 82102, 82103, 82104, 82105, 82106, 82107, 82108, 82109, 82110, 82112, 82116, 82118, 82120, 82121, 82123, 82125, 82126, 82128, 82129, 82130, 82131, 82132, 82133, 82138, 82142, 82143, 82144, 82146, 82148, 82201, 82202, 82203, 82236, 82238, 82240, 82241, 82250, 82251, 85005, 85006, 85007, 85008, 85009, 85010, 85011, 85014, 85025, 85026, 85027, 85028, 85029, 85031, 85032, 85033, 85040, 85041, 85042, 85043, 85044, 85046, 85047, 85049, 85050, 85066, 85067, 85068, 85069, 85080, 85081, 85082, 85083, 85084, 85086, 85088, 85090, 85096, 85103, 85104, 85105, 85106, 85107, 85108, 85110, 85117, 85141, 85143, 85144, 85145, 85149, 85164, 85165, 85167, 85168, 85169, 85170, 85171, 85172, 85174, 85175, 85176, 85177, 85178, 85179, 85180, 85213, 85214, 85215, 85216, 85217, 85218, 85219, 85238, 85239, 85243, 85244, 85245, 85246, 85247, 85248, 85249, 85250, 85252, 85253, 85254, 85255, 85257, 85260, 85296, 85299, 85300, 85302, 85305, 85306, 85307, 85308, 85309, 85312, 85315, 85316, 85317, 85319, 85320, 85322, 85329, 85330, 85332, 85333, 85334, 85335, 85337, 85338, 85355, 85373, 85475, 95001, 95007, 95008, 95009, 820001, 822002, 822003, 822004, 822005, 822009, 822017, 850004, 850015, 920003 & 950001.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAPinnacle3 Radiation Therapy Planning System, Model numbers 870231 and 870237.21 de septiembre de 2022
- FDABrilliance 64 Model # 728231, computed tomography x-ray system14 de agosto de 2019
- FDAIngenuity Core128 Model # 728323, computed tomography x-ray system14 de agosto de 2019
- FDABrightView XCT , Model Number 88248231 de julio de 2019
- FDAAlaris PC unit model 8015 with software version 9.12 Product Usage: The Alaris PC unit is the central24 de abril de 2013
- FDAIntegra XKnife® Stereotactic Radiosurgery and Radiotherapy Treatment Planning, Software Versions 5.0.1 and24 de abril de 2013
Preguntas sobre este retiro
¿Intellispace Portal (ISP) Philips Medical Systems (Cleveland), Inc., IntelliSpace Portal (ISP) is a multimodality (CT, NM, and MR) thin-client applications server that delivers full diagnostic viewing and clinical applications to the enterprise. IntelliSpace Portal (ISP) is a medical software system that allows multiple users to remotely access IntelliSpace Workspace Portal (ISP) from compatible computers on a network. The system allows networking, selection, processing and filming of multimodality DICOM images. Both the client and server software are only for use with off the shelf hardware technology that meets defined minimum specifications. trained physicians may use the images as a basis for diagnosis upon ensuring that monitor quality, ambient light conditions and image compression ratios are consistent with clinical application. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1102-2013 el 24 de abril de 2013. Peligro reportado: Philips was notified of a problem using the IntelliSpace Portal with software version 4.0.2 with the URL PACS integration. Philips was informed that when loading a study to an analysis application while using URL PACS integration configuration, after another study was opened by a review application, the Analysis monitor is updated with the study of the current patient, but the Review monitor conti
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter