ABX PENTRA Multical. ABX PENTRA N Control. ABX PENTRA P Control. The ABX PENTRA 400 and PC200 are benchtop clinical chemistry analyzers using two measuring principals: absorbance and ion selective electrodes. The ABX PENTRA Multical is a lyophilized human serum calibrator with chemical additives and materials of biological origin. The assigned values of the calibrator components are given in the enclosed annex, ensuring optimal calibration of the appropriate HORIBA ABX SAS methods on the ABX PENTRA Clinical Chemistry Analyzer. This calibrator is provided in ten vials of 3 ml.
Publicado el 12 de marzo de 2014 · Inicio del retiro: 20 de enero de 2014
Resumen
FDA registra el retiro Z-1146-2014 de dispositivos médicos de Clase II, publicado el 12 de marzo de 2014 e iniciado el 20 de enero de 2014. Empresa responsable: Horiba Instruments, Inc dba Horiba Medical; Estado: Finalizado; Cantidad reportada: 2,023 units; Lote: Model Numbers:Multical - A11A01652, N Control - A11A01653, P Control - A11A01654. Lot Numbers: MultiCal: 1204101 and 1304901, N Control: 1300801, P Control: 1300601. Expiration Date (s): MultiCal: 1204101 Expires on 05/31/14, MultiCal: 1304901 Expires on 07/15/15, N Control: 1300801 Expires on 11/05/14, P Control: 1300601 Expires on 11/05/14. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
HORIBA Medical is informing all ABX PENTRA MultiCal, N Control, and P Control customers running the Creatinine Rate Blank Method assay on the ABX PENTRA 400 and PC200 that the Target Values for Creatinine Rate Blank Method have been updated. This issue is only applicable to Creatinine Rate Blank Method values reported on the ABX PENTRA 400 and PC200 Chemistry Analyzers only. No other analyzers
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Horiba Instruments, Inc dba Horiba Medical
- Distribuidor
- United States nationwide distribution.
- Lote
- Model Numbers:Multical - A11A01652, N Control - A11A01653, P Control - A11A01654. Lot Numbers: MultiCal: 1204101 and 1304901, N Control: 1300801, P Control: 1300601. Expiration Date (s): MultiCal: 1204101 Expires on 05/31/14, MultiCal: 1304901 Expires on 07/15/15, N Control: 1300801 Expires on 11/05/14, P Control: 1300601 Expires on 11/05/14.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAABX Pentra Micro ALBUMIN 2 CP, Serum/Plasma, REF 1300032563 The device is intended for use in conjunction with4 de julio de 2018
- FDAABX PENTRA Magnesium RTU Reagent, Part No. A11A01646 Product Usage: ABX PENTRA Magnesium RTU reagent with1 de octubre de 2014
- FDADevice Name: ABX PENTRA ML Model/Part #(s): A11A01652 Lot/Serial #(s): N/A Software Version(s): All versions19 de marzo de 2014
- FDASimplexa Flu AlB & RSV Direct Assay Kits, intended for use as an aid in the differential diagnosis of12 de marzo de 2014
- FDAGE Healthcare Solar 8000M and Solar 8000i Patient Monitor software version 5.4, 5.5, or 5.6 with the12 de marzo de 2014
- FDAABX PENTRA N Control, Model No. A11A01653 The ABX PENTRA N Control is for use in quality control by monitoring12 de febrero de 2014
Preguntas sobre este retiro
¿ABX PENTRA Multical. ABX PENTRA N Control. ABX PENTRA P Control. The ABX PENTRA 400 and PC200 are benchtop clinical chemistry analyzers using two measuring principals: absorbance and ion selective electrodes. The ABX PENTRA Multical is a lyophilized human serum calibrator with chemical additives and materials of biological origin. The assigned values of the calibrator components are given in the enclosed annex, ensuring optimal calibration of the appropriate HORIBA ABX SAS methods on the ABX PENTRA Clinical Chemistry Analyzer. This calibrator is provided in ten vials of 3 ml. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1146-2014 el 12 de marzo de 2014. Peligro reportado: HORIBA Medical is informing all ABX PENTRA MultiCal, N Control, and P Control customers running the Creatinine Rate Blank Method assay on the ABX PENTRA 400 and PC200 that the Target Values for Creatinine Rate Blank Method have been updated. This issue is only applicable to Creatinine Rate Blank Method values reported on the ABX PENTRA 400 and PC200 Chemistry Analyzers only. No other analyzers
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter