Stryker Howmedica Osteonics Corp. AxSOS System Instrument. The AxSOS Locking Plate System is designed to treat periarticular or intraarticular fractures of the Distal Femur, Proximal Humerus, Proximal Tibia, and the Distal Tibia. Per the operative technique, " A 4.3 mm Drill (REF 702743) is then used to drill the core hole for the locking screw. Using a fluoroscopy, check the correct depth of the drill, and measure the length of the screw. The Drill Sleeve should now be removed, and the correct length 5.0 mm Locking Screw is inserted using the Screwdriver T20 and Screw Holding Sleeve."
Publicado el 8 de mayo de 2013 · Inicio del retiro: 28 de marzo de 2013
Resumen
FDA registra el retiro Z-1210-2013 de dispositivos médicos de Clase II, publicado el 8 de mayo de 2013 e iniciado el 28 de marzo de 2013. Empresa responsable: Stryker Howmedica Osteonics Corp.; Estado: Finalizado; Cantidad reportada: 107 units; Lote: Catalog number 703615. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Stryker Orthopaedics has received a report indicating that the scale of the AxSOS Calibrated Drill Bit 4.3,,X263mm, flat , manufactured by Stryker Osteosynthesis, is incorrect. The incorrect scale could result in the surgeon selecting a screw that is 10mm longer than intended.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Stryker Howmedica Osteonics Corp.
- Distribuidor
- Nationwide Distribution including the states of AL, AZ, IL, KY, LA, NJ, NC, FL, CA, NM, MO, TN, VA, CO, MA, TX, IN and MT.
- Lote
- Catalog number 703615
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAHoffmann LRF Safety Clip (Bone Transport Strut), 14 x 12 x 5 mm, REF 4933-0-401, NDC31 de mayo de 2017
- FDASupracondylar Nail T2 SCN 09x300 mm T2 Supracondylar Nail System is intended to provide strong and stable21 de septiembre de 2016
- FDAStryker Orthopaedics Specialty Triathlon Tibial Alignment Handle with Secondary Lock, Non-Sterile, Used to17 de agosto de 2016
- FDAStryker, T2 Guidewire, Ball-Tipped, 3 x 1000mm Product Usage: Intended for the temporary stabilization of bone2 de marzo de 2016
- FDAGE Healthcare, Optima MR360/Brivo MR355 (HDsv), GE 0.35T Signa Ovation HD MR System, GE 0.2T Signa Profile HD8 de mayo de 2013
- FDAMSK EXTREME MR SCANNER, MODEL AA5000 The intended for use as a diagnostic imaging device to produce axial,8 de mayo de 2013
Preguntas sobre este retiro
¿Stryker Howmedica Osteonics Corp. AxSOS System Instrument. The AxSOS Locking Plate System is designed to treat periarticular or intraarticular fractures of the Distal Femur, Proximal Humerus, Proximal Tibia, and the Distal Tibia. Per the operative technique, " A 4.3 mm Drill (REF 702743) is then used to drill the core hole for the locking screw. Using a fluoroscopy, check the correct depth of the drill, and measure the length of the screw. The Drill Sleeve should now be removed, and the correct length 5.0 mm Locking Screw is inserted using the Screwdriver T20 and Screw Holding Sleeve." fue retirado del mercado?
Sí. FDA publicó el retiro Z-1210-2013 el 8 de mayo de 2013. Peligro reportado: Stryker Orthopaedics has received a report indicating that the scale of the AxSOS Calibrated Drill Bit 4.3,,X263mm, flat , manufactured by Stryker Osteosynthesis, is incorrect. The incorrect scale could result in the surgeon selecting a screw that is 10mm longer than intended.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter